52
Nut
Spring washer
Plain washer
Bolt
Transmitter
Transmitter
Bracket
Plain washer
Spring washer
Mounting bolt
(M8 12)
Transmitter
Mesurement chamber cover
Bracket
Plain washer
Spring washer
Mounting bolt
(M8 12)
Pipe mounting
FKC, FKG and FKA models
FKD, FKB and FKM models
FKC, FKG and FKA models
FKD, FKB and FKM models
(1) Fasten the transmitter to a vertical or horizontal pipe using the supplied U-bolt (Tightening
torque approximately 15 Nm (1.5 kgfm)<11ft-lb>).
(2) Use a pipe of outside diameter ø60.5mm
Fasten tthe transmitter to the wall with M8 bolts.
Flange mounting
Position the remote seal of the transmitter in front of the counter
fl
ange. Fix the transmitter to the
fl
ange using the adapter boltings
Mount the bracket on the transmitter as shown below.
Bracket mounting
Mounting
- The transmitter is heavy. Be careful when handling it.
- The indicated installation and the wiring conditions of the transmitter must be
strictly followed.
- A wrong manipulation can be the reason of a disfunction of the transmitter.
- During installation, make sure that no part which could be the reason of the dis-
function or a danger is located inside the electronics housing.
- When the transmitter is installed in a
fl
ame proof area :
The local indicator position must not be changed
The position of the electronics housing must not be changed
- The isolating valves and manifold must correspond to the maximum pressure in
the pipe. If the valves and the connections do not correspond to the max pipe
pressure, there will be a risk of gas leakage or dangerous
fl
uid.
- The pipes need to be according the temperature and the pressure standards.
DANGER