Rerange (Set LRV/URV calibration)
Application to level measurement at
change of level (LRV/URV)
Functions of RERANGE can be made with
HHC Version 6.0 or upward.
When the lower range value (LRV) and
upper range value (URV) need to be ad-
justed again during measurement of tank
level, the measurement levels can be
changed at the same time by setting the
LRV or URV to be adjusted from FXW.
Apply an input pressure required for re-
range of LRV at display of 3 and press
ENT twice.
In this way, the rerange of LRV is com-
pleted, then the new measurement range
LRV and URV, which conforms to the ac-
tual input pressure, is displayed.
When rerange is made at a point other
than 0%, input the set value (PV%) of that
point at display of 3 , and press ENT at
display of 4 while applying a correspond-
ing pressure. In this way, the measure-
ment range can be changed to the input
corresponding to that pressure.
Apply an input pressure required for re-
range of URV at display of 6 and press
ENT twice. The rerange of URV is com-
pleted, then the new measurement range
LRV and URV corresponding to the actual
input pressure is displayed. When rerange
is made at a point other than 100%, input
the set value (PV%) of that point at display
of 6 and press ENT at display of 7 while
applying a corresponding pressure. In
this way, the measurement range can be
changed to the input corresponding to that
pressure.
Note)
The unit of LRV/URV at 5 and 8 are
displayed in the unit selected by Menu N°
4:UNIT.
I-2:
RERANGE
6
< ENT > < CL >
INC
2
3
4
1
CL
CHNG
CL
CL
LRV
URV
I:
RERANGE
< INC > < CHNG >
I-1:
RERANGE
< LRV > < URV > < CL >
URV 100.00%
CL
ENT
I-3:
RERANGE
CHANGE
CHNG OK? < ENT/CL >
LRV . %
7
CL
CL
CL
ENT
I-3:
RERANGE
CHANGE
CHNG OK? < ENT/CL >
URV
%
5
ENT
I-5: RERANGE
< CL >
LRV
p
kPa
LRV
p
kPa
1:
TAG N
o
8
ENT
I-5: RERANGE
< CL >
URV
p
kPa
LRV
p
kPa
I-2:
RERANGE
< ENT > < CL >
LRV 0.00%
Adjustment point setting condition
-1.00% LRV 100.00%
0.00% URV 110.00%
In the case that point is out of setting limit.
(Ex)
In case of the actual scale speci
fi
cation with
a digital indicator provided, if the range is
changed, indicator display might not match.
So, setting is required again in response to
the display in the digital indicator (G:XMTR
DISPLAY).
In case of the actual scale speci
fi
cation with
an analog indicator provided, if the range is
changed, the scale for indicator might not en-
sure exact reading.
When CHNG is pressed at display of q, the fol-
lowing is displayed.
This means that RERANGE can't be made
because MENU N° H:LINEARIZE is set in EF-
FECTIVE.
In this case, press the CL key and set in IN-
VALID on the panel of N° H: LINEARIZE.
1-1: RERANGE
Can’t proceed.
Set Linearize
invalid.
<CL>
1-2: RERANGE
LRV 100.01%
SETTING ERR <CL>