Chapter 12: Specifications
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
12-15
pF = pressure from filter
P
ven
= venous pressure
P
pHF
= pre-filter pressure
Pre-filter pressure
Measuring range: –50 mmHg to +750 mmHg
For haemoperfusion: –50 mmHg to +500 mmHg
Blood pump
Delivery rate: 10 ml/min to 500 ml/min
(MPS: 10 ml/min to 300 ml/min)
(Citrate anticoagulation: 10 ml/min to 200 ml/min)
(Paed. treatment: 10 ml/min to 100 ml/min)
Control accuracy: ±5%
Inner line diameter: 6.4 mm
System accuracy of the delivered blood volume: ±10% over the
complete duration of the treatment for typical treatment situations.
Time-based standstill monitoring as a safety system against blood loss
through coagulation.
Operator stops the blood pump with the
[STOP]
key; immediate visual
alarm, acoustic alarm after 2 minutes.
Air detector
Response threshold: drop of the fluid level, air bubbles, foam or
microfoam (blood / air mixture) over the entire blood flow range of
0 ml/min to 300 ml/min or 500 ml/min
Method: ultrasonic transmission
Heparin pump
Pump type: syringe pump
Delivery rate: 0.1 ml/h to 25 ml/h
Resolution: 0.1 ml/h
Accuracy: ±5% for a delivery rate of 1 ml/h to 25 ml/h and a
measurement time of 2 hours up to a counterpressure of 1.2 bars
(calibrated for 30 ml Fresenius heparin syringes). With delivery rates
< 1.0 ml/h, the tolerance may exceed the specified ±5% accuracy.
Bolus administration: 0.5 ml to 5 ml in increments of 0.1 ml (the
maximum bolus amount to be injected is preset to 5 ml. This parameter
can be set to smaller volumes in the System parameters screen).
Ci-Ca drop counter
Measuring range: 0 to 5 drops / second
Citrate pump
*
Pump type: roller pump
Delivery accuracy: pressure-dependent, ±10% (above typical pressure
range); for delivery rates < 6 ml/h, the deviation can be ±20%
Delivery rate: 10 to 600 ml/h (citrate / blood ratio: citrate concentration
per litre of delivered blood: 2–6 mmol/l)
Warning
Reduced treatment effectiveness
If the arterial pressure before the blood pump reaches extreme negative
values, the blood flow may be reduced, which will affect the
effectiveness of the treatment.
Содержание multiFiltrate
Страница 1: ...multiFiltrate Instructions for Use Software version 5 3 Edition 15A 2015 Part no M56 963 1 ...
Страница 2: ......
Страница 11: ...Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 xi 15 2 Instructions on the use of free software 15 2 ...
Страница 12: ...xii Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 16: ...Chapter 1 Index 1 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 42: ...Chapter 3 Design 3 10 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 166: ...Chapter 4 Operation 4 124 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 Interrupting the treatment ...
Страница 186: ...Chapter 5 Alarm processing 5 20 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 188: ...Chapter 6 Cleaning disinfection 6 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 218: ...Chapter 7 Functional description 7 30 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 228: ...Chapter 9 Installation 9 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 234: ...Chapter 11 Technical Safety Checks maintenance procedures 11 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 252: ...Chapter 12 Specifications 12 18 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 260: ...Chapter 13 Definitions 13 8 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 266: ...Chapter 14 Options 14 6 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 282: ...Chapter 15 Appendix 15 16 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...