Chapter 9: Installation
9-2
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
9.2
Installation / initial start-up requirements
When bringing the device from a cooler to a warmer room, allow approx.
2 hours for the device to adjust to the ambient temperature before
turning the unit on.
Charging the built-in
battery
On receipt of the device, charge the battery as follows:
– Use the power cable to connect the device to the power supply.
– Actuate the power switch to turn the device on.
– Leave the device on for 10 hours.
9.3
Important information on initial start-up
For initial start-up only
The following information is only intended for the initial start-up. This
information is not applicable for restarting devices that have been
removed from service or have temporarily been taken out of service.
Environmental conditions
Variations in temperature during transport may cause water
condensation on electrical parts. In the event of major variations in
temperature, allow sufficient time for the device to adjust to the ambient
temperature before start-up.
Qualification requirements
of testers
The initial system start-up must only be performed by the
manufacturer's service support organisation or a person authorised by
it.
The initial start-up must only be performed by personnel qualified to
perform the required procedures correctly based on their education,
training, knowledge, and experience. Furthermore, the personnel
performing the checks must be permitted to do so independently and
without outside interference.
Specifications
Any information on the specifications must be observed.
Documentation
The initial start-up report and further procedural information are
described in the Service Manual.
Reports are available on request.
The completion of the initial start-up must be entered in the Medical
Device Register.
Note
To reduce the risk of using the wrong citrate or calcium containers, only
one type of container (container size and concentration) should be used
throughout the hospital or a comparable organisation institution. Save
the same settings for citrate and calcium containers in the setup of all
devices of this organisation institution.
Содержание multiFiltrate
Страница 1: ...multiFiltrate Instructions for Use Software version 5 3 Edition 15A 2015 Part no M56 963 1 ...
Страница 2: ......
Страница 11: ...Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 xi 15 2 Instructions on the use of free software 15 2 ...
Страница 12: ...xii Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 16: ...Chapter 1 Index 1 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 42: ...Chapter 3 Design 3 10 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 166: ...Chapter 4 Operation 4 124 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 Interrupting the treatment ...
Страница 186: ...Chapter 5 Alarm processing 5 20 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 188: ...Chapter 6 Cleaning disinfection 6 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 218: ...Chapter 7 Functional description 7 30 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 228: ...Chapter 9 Installation 9 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 234: ...Chapter 11 Technical Safety Checks maintenance procedures 11 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 252: ...Chapter 12 Specifications 12 18 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 260: ...Chapter 13 Definitions 13 8 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 266: ...Chapter 14 Options 14 6 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 282: ...Chapter 15 Appendix 15 16 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...