Chapter 12: Specifications
12-14
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
12.11 Extracorporeal blood circuit and safety systems
Pressure transducer
(art. / ven. / pre-filter
pressure)
A filled pressure transducer is identified by the fact that no more
pressure fluctuation is detected at the pressure transducer within
5 minutes (arterial and pre-filter pressure) or 2 minutes (venous).
Arterial pressure
Display range: –280 mmHg to +300 mmHg
Resolution: 20 mmHg
Accuracy: 10 mmHg
Alarm window width: 20 mmHg to 200 mmHg, size and position
adjustable around actual pressure
Venous pressure
System safeguarding
against blood loss into the
environment
Display range: –80 mmHg to +500 mmHg
Resolution: 20 mmHg
Accuracy: 10 mmHg
Alarm window width: 20 mmHg to 200 mmHg, size and position
adjustable around actual pressure
Minimum alarm limits:
The optical detector senses non-opaque fluid: –20 mmHg during
treatment
The optical detector senses opaque fluid: 0 mmHg during treatment
The optical detector senses opaque fluid: –20 mmHg when
disconnecting
When the upper venous alarm limit is exceeded, the venous clamp will
not be closed to allow for release of the pressure in the system. In the
event of a simultaneous air detector alarm, the clamp will close.
TMP
Display range: –60 mmHg to +520 mmHg
Resolution: 20 mmHg
Accuracy: 10 mmHg
Alarm window width: 20 mmHg to 200 mmHg, size and position
adjustable around actual pressure
In MPS, the upper alarm limit is limited to 100 mmHg.
The TMP will be calculated using the following formula and will then be
displayed:
TMP = transmembrane pressure
Pven = venous pressure
P
pHF
= pre-filter pressure
P
Fil
= filtrate pressure
pF
Display range: –60 mmHg to +520 mmHg
Resolution: 20 mmHg
Accuracy: 10 mmHg
Alarm window width: 20 mmHg to 200 mmHg, size and position
adjustable around actual pressure
The pF is calculated using the following formula and then displayed:
pF = P
pHF
– P
ven
TMP
P
ven
P
PHF
+
2
--------------------------------- P
Fil
–
=
Содержание multiFiltrate
Страница 1: ...multiFiltrate Instructions for Use Software version 5 3 Edition 15A 2015 Part no M56 963 1 ...
Страница 2: ......
Страница 11: ...Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 xi 15 2 Instructions on the use of free software 15 2 ...
Страница 12: ...xii Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 16: ...Chapter 1 Index 1 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 42: ...Chapter 3 Design 3 10 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 166: ...Chapter 4 Operation 4 124 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 Interrupting the treatment ...
Страница 186: ...Chapter 5 Alarm processing 5 20 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 188: ...Chapter 6 Cleaning disinfection 6 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 218: ...Chapter 7 Functional description 7 30 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 228: ...Chapter 9 Installation 9 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 234: ...Chapter 11 Technical Safety Checks maintenance procedures 11 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 252: ...Chapter 12 Specifications 12 18 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 260: ...Chapter 13 Definitions 13 8 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 266: ...Chapter 14 Options 14 6 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 282: ...Chapter 15 Appendix 15 16 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...