Chapter 4: Operation
4-70
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Membrane plasma separation (MPS)
4.8.6.3
UF rinse
4.8.6.4
Recirculation / waiting for patient
Note
When using NaCl solutions with only one connector, make sure there is
enough NaCl solution.
After the rinse is completed and the
pre-defined volume used, the blood pump will
stop.
An audible signal will be given.
If using an NaCl solution with two
connectors:
Disconnect the venous patient line from
the empty bag and connect it to the NaCl
solution.
If using an NaCl solution with one
connector:
Do not change the existing connection.
Use the rotary selector to select
Start UF
rinse? [OK] to confirm!
and press
[OK]
.
Note
The filter life, which is shown in the status bar, is automatically added to
the service life of the tubing systems.
When the UF volume reaches 0 ml, the
extracorporeal blood circuit is in recirculation.
If using an NaCl solution with two
connectors:
Do not change the existing connection.
If using an NaCl solution with one
connector:
Stop recirculation by pressing the
[STOP]
key.
Press the [STOP] key for approx.
3 seconds.
Connect the venous patient line to the
arterial patient line using the recirculator.
Continue recirculation by pressing the
[START/RESET]
key.
Содержание multiFiltrate
Страница 1: ...multiFiltrate Instructions for Use Software version 5 3 Edition 15A 2015 Part no M56 963 1 ...
Страница 2: ......
Страница 11: ...Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 xi 15 2 Instructions on the use of free software 15 2 ...
Страница 12: ...xii Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 16: ...Chapter 1 Index 1 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 42: ...Chapter 3 Design 3 10 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 166: ...Chapter 4 Operation 4 124 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 Interrupting the treatment ...
Страница 186: ...Chapter 5 Alarm processing 5 20 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 188: ...Chapter 6 Cleaning disinfection 6 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 218: ...Chapter 7 Functional description 7 30 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 228: ...Chapter 9 Installation 9 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 234: ...Chapter 11 Technical Safety Checks maintenance procedures 11 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 252: ...Chapter 12 Specifications 12 18 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 260: ...Chapter 13 Definitions 13 8 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 266: ...Chapter 14 Options 14 6 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 282: ...Chapter 15 Appendix 15 16 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...