
Chapter 2: Important information
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
2-7
2.6.8
Interaction with other systems
The use of tube roller pumps may lead to minimal electrostatic
discharge into the tubing system due to friction on the pump segment.
As the charge is very low, these discharges do not represent a direct
hazard to patients or operators. If ECG units are used at the same time,
these discharges may, in rare cases, cause periodic interferences of the
ECG signal.
In order to minimise this interference, it is advisable to observe the
recommendations of the ECG device manufacturer, e.g.:
– Correct position of the electrodes
– Use of specific electrodes with low contact impedance
2.6.9
Restrictions
Regional anticoagulation
with citrate
Citrate anticoagulation is available for adult patients for CVVHD and
CVVHDF.
2.6.10 Risks of regional anticoagulation
If citrate anticoagulation is used in CRRT treatments, attention must be
paid to risks regarding the acid-base status and the electrolyte
concentrations. Further information (see chapter 7.3.2 on page 7-17).
Interrupting the treatment
Any interruption of the treatment (by temporarily disconnecting the
patient from the extracorporeal circuit) may give rise to additional risks.
This applies particularly if the blood is not retransfused to the patient
before the treatment is interrupted.
Retransfusion of the recirculated blood causes the risk of a high volume
of citrate being infused to the patient within a short time period. This
might, in particular, result in systemic hypocalcaemia. This risk is
especially high if you failed to deactivate balancing during the entire
recirculation phase.
2.6.11 Target group
The device may only be installed, operated, and used by individuals
with the appropriate training, knowledge, and experience and for whom
proof of instruction can be shown.
Содержание multiFiltrate
Страница 1: ...multiFiltrate Instructions for Use Software version 5 3 Edition 15A 2015 Part no M56 963 1 ...
Страница 2: ......
Страница 11: ...Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 xi 15 2 Instructions on the use of free software 15 2 ...
Страница 12: ...xii Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 16: ...Chapter 1 Index 1 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 42: ...Chapter 3 Design 3 10 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 166: ...Chapter 4 Operation 4 124 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 Interrupting the treatment ...
Страница 186: ...Chapter 5 Alarm processing 5 20 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 188: ...Chapter 6 Cleaning disinfection 6 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 218: ...Chapter 7 Functional description 7 30 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 228: ...Chapter 9 Installation 9 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 234: ...Chapter 11 Technical Safety Checks maintenance procedures 11 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 252: ...Chapter 12 Specifications 12 18 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 260: ...Chapter 13 Definitions 13 8 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 266: ...Chapter 14 Options 14 6 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 282: ...Chapter 15 Appendix 15 16 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...