4.11.8 Alarm limits menu
4.11.8.1 Preset alarm limits
Once treatment has started and the blood pump has achieved its target
delivery rate, the alarm limits are set after approx. 10 seconds, as
follows:
Arterial:
Centered with ±100 mmHg above / below actual pressure
Venous:
Lower alarm limit –20 mmHg below actual pressure
Upper alarm limit +100 mmHg above actual pressure
TMP:
Centered with ±100 mmHg above / below actual pressure
If the actual pressure value is near to the end of a scale, then the
pressure limits will be reduced accordingly. The absolute lower alarm
limit for venous pressure during treatment is +10 mmHg. The lower
pressure limit is adjusted accordingly.
Following this, the pressure limits should be checked by the operator
and adjusted individually as required.
Note
The system will only adopt a target and set change of the alarm limit
window width / alarm limit window position after confirmation with
[OK]
.
Note
If the message
"Setting alarm limits for venous pressure and TMP.
Please wait"
is displayed, the alarm limits menu is out of service. It is
possible to reset the alarm limits by pressing the
[START/RESET]
key.
Use the rotary selector to select the
required parameter and press
[OK]
.
For HP, the pressure from filter is shown with
the designation
pF
instead of
TMP
.
Содержание multiFiltrate
Страница 1: ...multiFiltrate Instructions for Use Software version 5 3 Edition 15A 2015 Part no M56 963 1 ...
Страница 2: ......
Страница 11: ...Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 xi 15 2 Instructions on the use of free software 15 2 ...
Страница 12: ...xii Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 16: ...Chapter 1 Index 1 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 42: ...Chapter 3 Design 3 10 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 166: ...Chapter 4 Operation 4 124 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 Interrupting the treatment ...
Страница 186: ...Chapter 5 Alarm processing 5 20 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 188: ...Chapter 6 Cleaning disinfection 6 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 218: ...Chapter 7 Functional description 7 30 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 228: ...Chapter 9 Installation 9 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 234: ...Chapter 11 Technical Safety Checks maintenance procedures 11 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 252: ...Chapter 12 Specifications 12 18 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 260: ...Chapter 13 Definitions 13 8 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 266: ...Chapter 14 Options 14 6 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 282: ...Chapter 15 Appendix 15 16 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...