Chapter 12: Specifications
12-8
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Voltage dips, short
interruptions, and
voltage variations
according to
IEC 61000-4-11
< 5% U
T
(> 95% dip
of the
U
T
) for
0.5 periods
< 5% U
T
(> 95% dip
of the
U
T
) for
0.5 periods
The quality of the supply voltage should
correspond to that of a typical commercial
and / or hospital environment. If the
operator of the multiFiltrate device requires
continuous operation, even in the event of
interruptions in the energy supply, it is
advisable to supply the multiFiltrate using
an uninterruptible power supply or a
battery.
40% U
T
(60% dip of
the
U
T
) for 5 periods
40% U
T
(60% dip of
the
U
T
) for 5 periods
70% U
T
(30% dip of
the
U
T
) for 25 periods
70% U
T
(30% dip of
the
U
T
) for 25 periods
< 5% U
T
(> 95% dip
of the
U
T
) for
5 seconds
< 5% U
T
(> 95% dip
of the
U
T
) for
5 seconds
Power frequency
magnetic fields
(50 / 60 Hz)
according to
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
correspond to the typical values found in
the commercial and hospital environment.
Note:
U
T
is the AC voltage before the test level is applied.
Portable and mobile radio devices should
not be any closer to the multiFiltrate device
(including the lines) than the
recommended safety distance, which is
calculated according to the relevant
equation for the transmission frequency.
Recommended safety distance:
Conducted RF
disturbances
according to
IEC 61000-4-6
3 V
eff
150 kHz to 80 MHz
3 V
eff
d = 1.2
√
P
For 150 kHz to < 80 MHz
Radiated RF
disturbances
according to
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
d = 1.2
√
P
For 80 MHz to < 800 MHz
d = 2.3
√
P
For 800 MHz to 2.5 GHz
Where P is the nominal power of the
transmitter in watts (W) in accordance with
the information from the transmitter
manufacturer and d is the recommended
safety distance in meters (m).
In accordance with an investigation on
site
a
, the field strength of stationary
transmitters should be lower than the
compliance level for all frequencies
b
.
Interference is possible in the
vicinity of devices which have this
symbol.
Содержание multiFiltrate
Страница 1: ...multiFiltrate Instructions for Use Software version 5 3 Edition 15A 2015 Part no M56 963 1 ...
Страница 2: ......
Страница 11: ...Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 xi 15 2 Instructions on the use of free software 15 2 ...
Страница 12: ...xii Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 16: ...Chapter 1 Index 1 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 42: ...Chapter 3 Design 3 10 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 166: ...Chapter 4 Operation 4 124 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 Interrupting the treatment ...
Страница 186: ...Chapter 5 Alarm processing 5 20 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 188: ...Chapter 6 Cleaning disinfection 6 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 218: ...Chapter 7 Functional description 7 30 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 228: ...Chapter 9 Installation 9 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 234: ...Chapter 11 Technical Safety Checks maintenance procedures 11 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 252: ...Chapter 12 Specifications 12 18 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 260: ...Chapter 13 Definitions 13 8 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 266: ...Chapter 14 Options 14 6 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 282: ...Chapter 15 Appendix 15 16 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...