Chapter 7: Functional description
7-12
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Frequently, an iso-oncotic human albumin solution is used. A lack of
coagulation factors or of other essential plasma components, whether
occurring as a result of the plasma exchange or otherwise, can be
counteracted by using fresh frozen plasma (FFP) as a replacement
solution, either completely or in part (in this case, preferably at the end
of the treatment).
In some cases, such as thrombotic thrombocytopenic purpura (TTP,
aka Moschcowitz syndrome), it is essential not only to remove the
pathologic plasma components, but also to reinfuse normal plasma
components with the replacement solution. Here, the use of FFP as the
replacement solution is recommended, or, alternatively, cryoprecipitate
plasma.
Efficacy
During MPS, typically one to two times the patient's plasma volume is
exchanged.
Due to the decrease in plasma concentration in the substances to be
removed in the course of the treatment, an MPS treatment ends as soon
as the prescribed plasma volume has been replaced. If medically
necessary, further MPS treatments will not be performed until the
following day at the earliest.
Treatment parameters
MPS
Min.
Max.
Resolution
Unit
Blood flow
10
300
10
ml/min
Plasma
Off / 10
10
1
ml/min
Temperature
Off / 35
37
0.5
°C
Cont. heparin adm.
Off / 0.1
25
0.1
ml/h
Содержание multiFiltrate
Страница 1: ...multiFiltrate Instructions for Use Software version 5 3 Edition 15A 2015 Part no M56 963 1 ...
Страница 2: ......
Страница 11: ...Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 xi 15 2 Instructions on the use of free software 15 2 ...
Страница 12: ...xii Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 16: ...Chapter 1 Index 1 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 42: ...Chapter 3 Design 3 10 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 166: ...Chapter 4 Operation 4 124 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 Interrupting the treatment ...
Страница 186: ...Chapter 5 Alarm processing 5 20 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 188: ...Chapter 6 Cleaning disinfection 6 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 218: ...Chapter 7 Functional description 7 30 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 228: ...Chapter 9 Installation 9 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 234: ...Chapter 11 Technical Safety Checks maintenance procedures 11 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 252: ...Chapter 12 Specifications 12 18 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 260: ...Chapter 13 Definitions 13 8 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 266: ...Chapter 14 Options 14 6 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 282: ...Chapter 15 Appendix 15 16 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...