![Fresenius Medical Care multiFiltrate Скачать руководство пользователя страница 147](http://html.mh-extra.com/html/fresenius-medical-care/multifiltrate/multifiltrate_instructions-for-use-manual_2331671147.webp)
Chapter 4: Operation
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
4-105
Treatment menu
4.11.2 Selecting Ci-Ca anticoagulation
Use the rotary selector to select
Yes
from
the menu field and press
[OK]
.
Use the rotary selector to select
All
conditions fulfilled? [OK] to confirm!
and press
[OK]
.
After deselecting citrate anticoagulation, the
multiFiltrate immediately switches to the
menu for dialysate bag change.
Warning
Ci-Ca CVVHD and Ci-Ca postCVVHDF treatments may only be
performed if calcium-free dialysate is used.
Warning
When starting the treatment, define the post-filter calcium value. If the
ionised calcium has not decreased at this point, it is absolutely
necessary that the tubing system and the solutions used are checked.
Содержание multiFiltrate
Страница 1: ...multiFiltrate Instructions for Use Software version 5 3 Edition 15A 2015 Part no M56 963 1 ...
Страница 2: ......
Страница 11: ...Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 xi 15 2 Instructions on the use of free software 15 2 ...
Страница 12: ...xii Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 16: ...Chapter 1 Index 1 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 42: ...Chapter 3 Design 3 10 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 166: ...Chapter 4 Operation 4 124 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 Interrupting the treatment ...
Страница 186: ...Chapter 5 Alarm processing 5 20 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 188: ...Chapter 6 Cleaning disinfection 6 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 218: ...Chapter 7 Functional description 7 30 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 228: ...Chapter 9 Installation 9 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 234: ...Chapter 11 Technical Safety Checks maintenance procedures 11 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 252: ...Chapter 12 Specifications 12 18 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 260: ...Chapter 13 Definitions 13 8 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 266: ...Chapter 14 Options 14 6 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 282: ...Chapter 15 Appendix 15 16 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...