Chapter 5: Alarm processing
5-4
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
The following alarm conditions are assigned high priority in the "New"
alarm scheme:
– Lower return pressure:
Potentially, the system may become disconnected, resulting in
ongoing blood loss via the access.
– Failure to detect the pump clips on the Ci-Ca pumps (no occlusion):
Potentially, the patient may suffer air infusion or blood loss via the
Ci-Ca tubing system.
In addition, the following conditions have an elevated risk and require
intensive observation and monitoring:
– Override condition following blood leak.
5.4
Monitoring function suppression
Alarm override /
confirmation
Suppresses the pressure alarms (alarm limits widened to the end of the
scale) for the time the blood pump starts running plus 10 seconds.
Overrides a blood leak alarm for around 2 minutes (for MPS, the blood
leak detector can also be deactivated in the
Treatment
menu).
It is not possible to override the air alarm.
5.5
Alarm system
Pressure monitoring
To avoid unnecessary false alarms, the monitoring window can be
temporarily extended, disabled, or its levels adjusted to the current
pressure following changes to relevant parameters, pressure alarms, or
issuing pump stop / start commands. These conditions are time-limited.
Monitoring at the display range thresholds remains unaffected.
Air infusion
To ensure that the stringent limit values for detecting air infusion are
always maintained, you may need to restrict the maximum blood flow
for low-weight patients (see chapter 12.11 on page 12-14).
Warning
When the monitoring function is suppressed, the operator is responsible
for the patient's safety.
Note
Air infusion alarm limits are dependent on blood flow and patient weight.
Full sensitivity at max. blood flow is achieved with patient weights
upwards of 45 kg.
If return pressure is negative (e.g., when deploying central venous
catheters) then the risk of an air infusion is present, even downstream
of the air detector.
Содержание multiFiltrate
Страница 1: ...multiFiltrate Instructions for Use Software version 5 3 Edition 15A 2015 Part no M56 963 1 ...
Страница 2: ......
Страница 11: ...Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 xi 15 2 Instructions on the use of free software 15 2 ...
Страница 12: ...xii Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 16: ...Chapter 1 Index 1 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 42: ...Chapter 3 Design 3 10 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 166: ...Chapter 4 Operation 4 124 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 Interrupting the treatment ...
Страница 186: ...Chapter 5 Alarm processing 5 20 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 188: ...Chapter 6 Cleaning disinfection 6 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 218: ...Chapter 7 Functional description 7 30 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 228: ...Chapter 9 Installation 9 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 234: ...Chapter 11 Technical Safety Checks maintenance procedures 11 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 252: ...Chapter 12 Specifications 12 18 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 260: ...Chapter 13 Definitions 13 8 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 266: ...Chapter 14 Options 14 6 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 282: ...Chapter 15 Appendix 15 16 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...