Chapter 9: Installation
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
9-3
9.4
Electrical installation
Power supply system
The national standards and regulations must be observed when
connecting the device to the power supply system.
Electromagnetic
compatibility
Please observe the following during installation and start-up:
(see chapter 12.5 on page 12-5)
Protective earth conductor
When using protection class I devices, the quality of the protective earth
conductor of the installation is of particular importance.
Power cable
If the power cable needs to be replaced, use only the original power
cable listed in the spare parts catalog. The use of additional extension
cables or multiway sockets / connectors is prohibited.
Equipotential bonding
Using the original accessories, connect the equipotential bonding
conductor to the rear of the device if this is required by law or for special
applications at the place of installation.
Leakage currents
If additional equipment not listed in the Accessories chapter is
connected to the device, there is a danger that the permitted leakage
currents will be exceeded.
Warning
The "type of protection against electric shock" for this device is
"Protection class I". To avoid the risk of electric shock, this device may
only be connected to a power supply network with a protective earth
conductor.
It must be taken into consideration that, in many countries, specific
regulations of the national authorities are in force.
Содержание multiFiltrate
Страница 1: ...multiFiltrate Instructions for Use Software version 5 3 Edition 15A 2015 Part no M56 963 1 ...
Страница 2: ......
Страница 11: ...Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 xi 15 2 Instructions on the use of free software 15 2 ...
Страница 12: ...xii Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 16: ...Chapter 1 Index 1 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 42: ...Chapter 3 Design 3 10 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 166: ...Chapter 4 Operation 4 124 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 Interrupting the treatment ...
Страница 186: ...Chapter 5 Alarm processing 5 20 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 188: ...Chapter 6 Cleaning disinfection 6 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 218: ...Chapter 7 Functional description 7 30 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 228: ...Chapter 9 Installation 9 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 234: ...Chapter 11 Technical Safety Checks maintenance procedures 11 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 252: ...Chapter 12 Specifications 12 18 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 260: ...Chapter 13 Definitions 13 8 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 266: ...Chapter 14 Options 14 6 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 282: ...Chapter 15 Appendix 15 16 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...