88
va06.0_0031g_Eng.fm
Inte
rventio
n pr
ocedu
re
s
Dismantling
When working with electronic components, we recommend wearing an anti-static bracelet
connected to earth and working on a foam anti-static mat.
Remove the angle bracket (marker 2) being careful not to pull out the CPU board flat
cable.
Disconnect the flat cable connecting the CPU board to the power supply at J1.
Pull out the RJ45 stopper (marker 3).
Unscrew the 2 screws (marker 4) holding the upper case (marker 5) to the base.
4
2
J1
5
4
3
Remove the upper case (marker 5), being careful not to pull out the display board flat
cable.
Disconnect the flat cable connecting the CPU board to the display board at J2.
Содержание Volumat Agilia
Страница 1: ...0 9 C E S Volumat Agilia Technical Manual ...
Страница 6: ...66 TM_Volumat_Agilia_Eng_TDM fm ...
Страница 8: ...8 va01_001e_Eng fm Introduction 1 2 Operation diagram ...
Страница 12: ...12 va01_001e_Eng fm Introduction ...
Страница 26: ...26 va02_001e_Eng fm Description and operation ...
Страница 30: ...30 va03 1_001e_Eng fm Description of the menus ...
Страница 62: ...62 va04 4_001e_Eng fm Preventive maintenance ...
Страница 70: ...70 va05 1_001e_Eng fm Troubleshooting ...
Страница 76: ...76 va06 0_001g_Eng fm Intervention procedures ...
Страница 82: ...82 va06 0_0011g_Eng fm Intervention procedures ...
Страница 86: ...86 va06 0_0021g_Eng fm Intervention procedures ...
Страница 94: ...94 va06 0_0040g_Eng fm Fiches d intervention ...
Страница 104: ...104 va06 0_0060g_Eng fm Fiches d intervention ...
Страница 110: ...110 va06 0_0070g_Eng fm Fiches d intervention ...
Страница 115: ...va06 0_0080g_Eng fm 115 Fiches d intervention 12 Remove the left flange marker 12 Pull out the pumping system ...
Страница 134: ...134 va07 1_001g_Eng fm Calibrations ...
Страница 136: ...136 va08 0_001d_Eng fm Spare parts catalogue ...