va04.4_001e_Eng.fm
59
Pr
ev
en
tive
m
a
int
en
a
nce
Fill the test administration set or the beaker with some water
ensuring the canula is dipped in the water (> 1cm).
Add several drops of oil to create a greasy film on the
surface of the water. This way the user will avoid any
measurement error due to evaporation of the water.
Place the test tube or the beaker in the centre of the scale
platform ensuring is not in touch with the scale.
Place the canula inside the test tube on the beaker ensuring
that it is dipped in the water.
The infusion line (needle/catheter extension) must not rest
on the scale/test tube unit.
Press "ON" (device on mains power mode).
Prime the infusion line using the "BOLUS" key.
Check for the absence of air bubbles.
Operating mode
Carry out a flow rate selection.
For low flow rates (< 5ml/h), confirm or wait 1 hour for infusion to stabilise. For high flow
rates, wait 10 to 30 minutes after the end of the infusion.
Carry out the scale taring:
00.00
g.
Start the infusion by pressing "VALIDATION", and start the
chronometer (if necessary, note the chronometer value).
After one hour, press "STOP" to stop the test.
Note the value (in grams) of the "infused" liquid.
Calculate the difference between the theoretical value and
the real value.
1gram = 1ml.
The error percentage is calculated according to the following
formula:
Measured value Theoretical value
–
Theoretical value
---------------------------------------------------------------------------------------------------
100
Error percentage
=
Содержание Volumat Agilia
Страница 1: ...0 9 C E S Volumat Agilia Technical Manual ...
Страница 6: ...66 TM_Volumat_Agilia_Eng_TDM fm ...
Страница 8: ...8 va01_001e_Eng fm Introduction 1 2 Operation diagram ...
Страница 12: ...12 va01_001e_Eng fm Introduction ...
Страница 26: ...26 va02_001e_Eng fm Description and operation ...
Страница 30: ...30 va03 1_001e_Eng fm Description of the menus ...
Страница 62: ...62 va04 4_001e_Eng fm Preventive maintenance ...
Страница 70: ...70 va05 1_001e_Eng fm Troubleshooting ...
Страница 76: ...76 va06 0_001g_Eng fm Intervention procedures ...
Страница 82: ...82 va06 0_0011g_Eng fm Intervention procedures ...
Страница 86: ...86 va06 0_0021g_Eng fm Intervention procedures ...
Страница 94: ...94 va06 0_0040g_Eng fm Fiches d intervention ...
Страница 104: ...104 va06 0_0060g_Eng fm Fiches d intervention ...
Страница 110: ...110 va06 0_0070g_Eng fm Fiches d intervention ...
Страница 115: ...va06 0_0080g_Eng fm 115 Fiches d intervention 12 Remove the left flange marker 12 Pull out the pumping system ...
Страница 134: ...134 va07 1_001g_Eng fm Calibrations ...
Страница 136: ...136 va08 0_001d_Eng fm Spare parts catalogue ...