56
va04.4_001e_Eng.fm
Pr
even
tive ma
inten
ance
Installation
Following the diagram below:
Anti-vibration measuring table
1 cm
Catheter extension bracket support
cable
RS 232
50 cm
Observe the equipment horizontal installation plan.
Fill the container with 600ml or more of distilled water.
Connect an administration set with the container.
Connect the canula with the distal end
of the administration set.
Fit the administration set on to the pump.
Fill the administration set with water. Take care that
no air is in the administration set before the measurement
is started.
Fill the test tube or the beaker with some water ensuring
the canula is dipped in the water (> 1cm).
Add several drops of oil to create a greasy film on the
surface of the water. This way the user will avoid any
measurement error due to evaporation of the water.
Place the test tube or the beaker in the centre of the scale
platform ensuring is not in touch with the scale.
Place the canula inside the test tube or the beaker ensuring
that it is dipped in the water.
The infusion line (needle/catheter extension) must not rest on the scale/test tube
unit.
Press "ON" (device on mains power mode).
Prime the infusion line using the "BOLUS" key.
Check for the absence of air bubbles.
Содержание Volumat Agilia
Страница 1: ...0 9 C E S Volumat Agilia Technical Manual ...
Страница 6: ...66 TM_Volumat_Agilia_Eng_TDM fm ...
Страница 8: ...8 va01_001e_Eng fm Introduction 1 2 Operation diagram ...
Страница 12: ...12 va01_001e_Eng fm Introduction ...
Страница 26: ...26 va02_001e_Eng fm Description and operation ...
Страница 30: ...30 va03 1_001e_Eng fm Description of the menus ...
Страница 62: ...62 va04 4_001e_Eng fm Preventive maintenance ...
Страница 70: ...70 va05 1_001e_Eng fm Troubleshooting ...
Страница 76: ...76 va06 0_001g_Eng fm Intervention procedures ...
Страница 82: ...82 va06 0_0011g_Eng fm Intervention procedures ...
Страница 86: ...86 va06 0_0021g_Eng fm Intervention procedures ...
Страница 94: ...94 va06 0_0040g_Eng fm Fiches d intervention ...
Страница 104: ...104 va06 0_0060g_Eng fm Fiches d intervention ...
Страница 110: ...110 va06 0_0070g_Eng fm Fiches d intervention ...
Страница 115: ...va06 0_0080g_Eng fm 115 Fiches d intervention 12 Remove the left flange marker 12 Pull out the pumping system ...
Страница 134: ...134 va07 1_001g_Eng fm Calibrations ...
Страница 136: ...136 va08 0_001d_Eng fm Spare parts catalogue ...