42
va04.3_001g_Eng.fm
Ma
in
te
na
nc
e pr
év
en
tive
4.4.13 Test 11, Ageing test
This text checks the motor block/pumping system is functioning correctly as well as the pump
occlusivity.
Test 11
, press "enter".
The main, "
Ageing test
" menu displays.
When the pump is fitted with a tube, open the
administration set clamp by pressing the bolus key.
Select your submenu using the "OK" key.
Flow rate
: between 0.1 and 1200ml/h
Modify the value using the following selection keys:
for the hundreds and tens, use the double arrows to
increase or decrease
for the units, increase or decrease using the single
arrows.
Press "OK" to validate and to move to the next step.
Mode
: Go forward or synchro or go forwards/backwards
or backwards
Modify your choice using the single arrow selection keys
.
Press "OK" to confirm and move to the next step.
Steps numbers
or turns
number
:
Content or value lying between 1 and 9999
Modify the value using the single arrow selection keys
Press "OK" to confirm and move to the next step.
Current
: high, low, auto or error.
Modify the value using the single arrow selection keys
,
Press "OK" to confirm and move to the next step.
Press "OK" to confirm and start up the test.
The "
Ageing test
" menu displays:
The flow rate and current category for the
motor command are displayed at the top
of the screen.
The number of steps decreases according to
the actual change for the number of steps.
Press "exit" to return to the "
Test
" menu.
During the test, the rotation control is activated and the device set in error if a problem is
detected.
In a test with a defined number of steps/cycles or using a synchro mode, the test ends
when the requested position is reached.
Содержание Volumat Agilia
Страница 1: ...0 9 C E S Volumat Agilia Technical Manual ...
Страница 6: ...66 TM_Volumat_Agilia_Eng_TDM fm ...
Страница 8: ...8 va01_001e_Eng fm Introduction 1 2 Operation diagram ...
Страница 12: ...12 va01_001e_Eng fm Introduction ...
Страница 26: ...26 va02_001e_Eng fm Description and operation ...
Страница 30: ...30 va03 1_001e_Eng fm Description of the menus ...
Страница 62: ...62 va04 4_001e_Eng fm Preventive maintenance ...
Страница 70: ...70 va05 1_001e_Eng fm Troubleshooting ...
Страница 76: ...76 va06 0_001g_Eng fm Intervention procedures ...
Страница 82: ...82 va06 0_0011g_Eng fm Intervention procedures ...
Страница 86: ...86 va06 0_0021g_Eng fm Intervention procedures ...
Страница 94: ...94 va06 0_0040g_Eng fm Fiches d intervention ...
Страница 104: ...104 va06 0_0060g_Eng fm Fiches d intervention ...
Страница 110: ...110 va06 0_0070g_Eng fm Fiches d intervention ...
Страница 115: ...va06 0_0080g_Eng fm 115 Fiches d intervention 12 Remove the left flange marker 12 Pull out the pumping system ...
Страница 134: ...134 va07 1_001g_Eng fm Calibrations ...
Страница 136: ...136 va08 0_001d_Eng fm Spare parts catalogue ...