J10 connector to opto motor
Pin
Description
1
GND
Ground
2
MP_OPTO/
Motor opto rotation state
3
MP_CD_OPTO
Opto power supply
4
GND
Opto power supply return
J11 connector to display board
Pin
Description
A1
GND
Ground
B1
VBAT
Battery power supply or + "10V mains"
A2
+3V3-
PERM_PROT
Diagnostic signal (correct boards connection)
B2
-
Not connected
A3
GND
Ground
B3
+3V3
Power 3V3
A4
TON/OFF
Keyboard ON/OFF key
B4
MISO
SPI bus MISO signal
A5
GND
Ground
B5
MOSI
SPI bus MOSI signal
A6
GND
Ground
B6
CLK
SPI bus clock signal
A7
GND
Ground
B7
RESET/
Control signal
A8
RSLCD
Control signal (intended for LCD display)
B8
CSLCD/
Control signal
A9
CSLDCD/
Control signal
B9
CD_BUZ_AFF
Buzzer command
A10
GND
Ground
B10
S_ANA_DOOR
Analogue signal sent by the "DOOR" sensor
A11
+3V3
Power 3.3V
B11
GND
Ground
A12
-
Not connected
B12
PRES_CARTES
Diagnostic signal (correct board connection)
A13
VBAT
Battery power supply or "+10V mains"
B13
GND
Ground
J12 connector to drop detector
Pin
Description
1
GND
Ground
2
PULSE_goutte
Sensor return signal
3
GND
Ground
4
GND
Ground
5
CDLED_goutte
Sensor power supply
6
GND
Ground
18
va02_001e_Eng.fm
Descrip
tion an
d op
era
tion
Содержание Volumat Agilia
Страница 1: ...0 9 C E S Volumat Agilia Technical Manual ...
Страница 6: ...66 TM_Volumat_Agilia_Eng_TDM fm ...
Страница 8: ...8 va01_001e_Eng fm Introduction 1 2 Operation diagram ...
Страница 12: ...12 va01_001e_Eng fm Introduction ...
Страница 26: ...26 va02_001e_Eng fm Description and operation ...
Страница 30: ...30 va03 1_001e_Eng fm Description of the menus ...
Страница 62: ...62 va04 4_001e_Eng fm Preventive maintenance ...
Страница 70: ...70 va05 1_001e_Eng fm Troubleshooting ...
Страница 76: ...76 va06 0_001g_Eng fm Intervention procedures ...
Страница 82: ...82 va06 0_0011g_Eng fm Intervention procedures ...
Страница 86: ...86 va06 0_0021g_Eng fm Intervention procedures ...
Страница 94: ...94 va06 0_0040g_Eng fm Fiches d intervention ...
Страница 104: ...104 va06 0_0060g_Eng fm Fiches d intervention ...
Страница 110: ...110 va06 0_0070g_Eng fm Fiches d intervention ...
Страница 115: ...va06 0_0080g_Eng fm 115 Fiches d intervention 12 Remove the left flange marker 12 Pull out the pumping system ...
Страница 134: ...134 va07 1_001g_Eng fm Calibrations ...
Страница 136: ...136 va08 0_001d_Eng fm Spare parts catalogue ...