va06.0_0090g_Eng.fm
119
Fiche
s d’in
te
rven
tion
Disconnect the flat cable leading from the CPU board to the display board at J2.
Disconnect the flat cables leading to the CPU board at:
J2 and J13
J7, J3 and J10
J4 and J12
J1
Unscrew the holding screws (marker 6) holding the CPU board to the flange.
J12
6
J1
J3
J2
J13
J4
J10
J7
Remove the whole CPU board being careful not to pull out the flat cable on the inside.
Содержание Volumat Agilia
Страница 1: ...0 9 C E S Volumat Agilia Technical Manual ...
Страница 6: ...66 TM_Volumat_Agilia_Eng_TDM fm ...
Страница 8: ...8 va01_001e_Eng fm Introduction 1 2 Operation diagram ...
Страница 12: ...12 va01_001e_Eng fm Introduction ...
Страница 26: ...26 va02_001e_Eng fm Description and operation ...
Страница 30: ...30 va03 1_001e_Eng fm Description of the menus ...
Страница 62: ...62 va04 4_001e_Eng fm Preventive maintenance ...
Страница 70: ...70 va05 1_001e_Eng fm Troubleshooting ...
Страница 76: ...76 va06 0_001g_Eng fm Intervention procedures ...
Страница 82: ...82 va06 0_0011g_Eng fm Intervention procedures ...
Страница 86: ...86 va06 0_0021g_Eng fm Intervention procedures ...
Страница 94: ...94 va06 0_0040g_Eng fm Fiches d intervention ...
Страница 104: ...104 va06 0_0060g_Eng fm Fiches d intervention ...
Страница 110: ...110 va06 0_0070g_Eng fm Fiches d intervention ...
Страница 115: ...va06 0_0080g_Eng fm 115 Fiches d intervention 12 Remove the left flange marker 12 Pull out the pumping system ...
Страница 134: ...134 va07 1_001g_Eng fm Calibrations ...
Страница 136: ...136 va08 0_001d_Eng fm Spare parts catalogue ...