Problem
Cause
Recommended action
Flow rate
variance too great
A mechanical part of the
pumping mechanism or the
CPU board has been
changed without the
calibration of the infusion
corrector coefficient being
carried out
Carry out a flow rate test,
calibrate the infusion corrector
coefficient
Redo the test to ensure that the
correction is right
Display default:
segments, LED
LED and/or connectors
defective
Check the general state of the
display board
Check the contrast
Check the connectors
Replace the display board
The device
switches off when
it is disconnected
from the mains
Battery is completely
discharged
Check the battery connection
Charge the battery
Replace the battery
The mains
indicator does not
light up
Mains fuses are out of
order
Replace the mains fuses
Check that the mains connector
has been correctly connected
Replace the power supply board
Battery alarm
while it has been
correctly charged
Battery is completely
discharged
Check the charging voltage
Replace the battery
The device
switches on or off
on its own
Defective keyboard
Defective CPU board
Check and replace the keyboard
if necessary
Replace the CPU board and
reconfigure completely
The keyboard
keys do not work
Defective keyboard
Check and replace the keyboard
if necessary
Mechanical
elements are
damaged
The pump has been
dropped
Do not use the device.
Contact your biomedical
department or our Technical
Service.
va05.1_001e_Eng.fm
69
Tr
oub
lesho
o
ting
Содержание Volumat Agilia
Страница 1: ...0 9 C E S Volumat Agilia Technical Manual ...
Страница 6: ...66 TM_Volumat_Agilia_Eng_TDM fm ...
Страница 8: ...8 va01_001e_Eng fm Introduction 1 2 Operation diagram ...
Страница 12: ...12 va01_001e_Eng fm Introduction ...
Страница 26: ...26 va02_001e_Eng fm Description and operation ...
Страница 30: ...30 va03 1_001e_Eng fm Description of the menus ...
Страница 62: ...62 va04 4_001e_Eng fm Preventive maintenance ...
Страница 70: ...70 va05 1_001e_Eng fm Troubleshooting ...
Страница 76: ...76 va06 0_001g_Eng fm Intervention procedures ...
Страница 82: ...82 va06 0_0011g_Eng fm Intervention procedures ...
Страница 86: ...86 va06 0_0021g_Eng fm Intervention procedures ...
Страница 94: ...94 va06 0_0040g_Eng fm Fiches d intervention ...
Страница 104: ...104 va06 0_0060g_Eng fm Fiches d intervention ...
Страница 110: ...110 va06 0_0070g_Eng fm Fiches d intervention ...
Страница 115: ...va06 0_0080g_Eng fm 115 Fiches d intervention 12 Remove the left flange marker 12 Pull out the pumping system ...
Страница 134: ...134 va07 1_001g_Eng fm Calibrations ...
Страница 136: ...136 va08 0_001d_Eng fm Spare parts catalogue ...