74
va05.2_001g_Eng.fm
Dia
gn
os
tic
41
Upstream pressure sensor
out of range
Select test 5.
Check the change in the state of the
sensor by applying pressure to the sensor.
Check the connections.
Calibrate the upstream pressure sensor.
42
Downstream pressure
sensor out of range
Select test 6.
Check the change in state of the sensor by
applying a pressure on the sensor.
Check the connections
Calibrate the downstream pressure
sensor.
99
Activation of the watch dog
Replace the power supply board to check
the functionality of the secondary µC.
Replace the CPU board to check the
functionality of the main µP.
Carry out a complete configuration, when
replacing the CPU board, using the
maintenance software.
Error
Code
Description
Recommended action
Содержание Volumat Agilia
Страница 1: ...0 9 C E S Volumat Agilia Technical Manual ...
Страница 6: ...66 TM_Volumat_Agilia_Eng_TDM fm ...
Страница 8: ...8 va01_001e_Eng fm Introduction 1 2 Operation diagram ...
Страница 12: ...12 va01_001e_Eng fm Introduction ...
Страница 26: ...26 va02_001e_Eng fm Description and operation ...
Страница 30: ...30 va03 1_001e_Eng fm Description of the menus ...
Страница 62: ...62 va04 4_001e_Eng fm Preventive maintenance ...
Страница 70: ...70 va05 1_001e_Eng fm Troubleshooting ...
Страница 76: ...76 va06 0_001g_Eng fm Intervention procedures ...
Страница 82: ...82 va06 0_0011g_Eng fm Intervention procedures ...
Страница 86: ...86 va06 0_0021g_Eng fm Intervention procedures ...
Страница 94: ...94 va06 0_0040g_Eng fm Fiches d intervention ...
Страница 104: ...104 va06 0_0060g_Eng fm Fiches d intervention ...
Страница 110: ...110 va06 0_0070g_Eng fm Fiches d intervention ...
Страница 115: ...va06 0_0080g_Eng fm 115 Fiches d intervention 12 Remove the left flange marker 12 Pull out the pumping system ...
Страница 134: ...134 va07 1_001g_Eng fm Calibrations ...
Страница 136: ...136 va08 0_001d_Eng fm Spare parts catalogue ...