4.4.22 Checking the presence limit alarm
To carry out this operation, exit the test mode and press "OFF".
Carry out the following assembly
T
1 m
Pump
50 cm
Container
3-way stopcock
Manometer
Bag
:
Press the "ON/OFF"key.
Place the administration set on the device and close to
the door.
Place the manometer at the tube outlet.
Select a flow rate of 125 ml/h.
Adjust the pressure threshold to 750mmHg.
Open the 3-way stopcok to a container.
Start the infusion and leave it running for 30 seconds.
Position the 3-way stopcok on the manometer side.
Check that the alarm is triggered for a value of 1bar
± 0.15 bar.
Stop the infusion cycle by pressing "STOP".
When the value of the measured pressure, lies outside of the limit selected, recalibrate
the pressure sensors (see "
Etal 1
Upstream and downstream pressure sensors
calibration").
4.4.23 Mains/battery operation check
To carry out this procedure, exit the test mode pressing the "OFF" key.
Connect the device to the mains.
Check the functioning of mains presence LED (plug-
shaped indicator).
Disconnect the device from the mains.
Check that the mains indicator goes out.
Содержание Volumat Agilia
Страница 1: ...0 9 C E S Volumat Agilia Technical Manual ...
Страница 6: ...66 TM_Volumat_Agilia_Eng_TDM fm ...
Страница 8: ...8 va01_001e_Eng fm Introduction 1 2 Operation diagram ...
Страница 12: ...12 va01_001e_Eng fm Introduction ...
Страница 26: ...26 va02_001e_Eng fm Description and operation ...
Страница 30: ...30 va03 1_001e_Eng fm Description of the menus ...
Страница 62: ...62 va04 4_001e_Eng fm Preventive maintenance ...
Страница 70: ...70 va05 1_001e_Eng fm Troubleshooting ...
Страница 76: ...76 va06 0_001g_Eng fm Intervention procedures ...
Страница 82: ...82 va06 0_0011g_Eng fm Intervention procedures ...
Страница 86: ...86 va06 0_0021g_Eng fm Intervention procedures ...
Страница 94: ...94 va06 0_0040g_Eng fm Fiches d intervention ...
Страница 104: ...104 va06 0_0060g_Eng fm Fiches d intervention ...
Страница 110: ...110 va06 0_0070g_Eng fm Fiches d intervention ...
Страница 115: ...va06 0_0080g_Eng fm 115 Fiches d intervention 12 Remove the left flange marker 12 Pull out the pumping system ...
Страница 134: ...134 va07 1_001g_Eng fm Calibrations ...
Страница 136: ...136 va08 0_001d_Eng fm Spare parts catalogue ...