even
tive ma
inten
ance
4.9 Use of the internal battery
This device is provided with Li-Ion battery. When the device is disconnected from the mains, it
automatically switches to battery mode.
Before starting for the first time, charge the battery for approx. 5 hours by connecting the
power supply cord without using the device.
If possible, if not using device for an extended period (higher than 2 months), remove it from
the device and store it in a place according to recommended use conditions. If not possible to
remove battery or for a short period (lower than 2 months), it is recommended to charge the
battery at least once a month, by leaving the device connected on mains power supply 8
hours at least (device OFF).
You can also recharge a Lithium-ion battery whenever convenient, without the full charge /
discharge cycle necessary to get full capacity when fully charged.
In order to maximize the life time and the performances of the battery:
Use and store in a cool place.
In use, leave the device connected to mains power supply to maintain the charge of the
battery and the maximum capacity when possible.
Lithium Ion rechargeable battery, handle with care !
Do not incinerate or place near an open flame.
Do not drop, crush, puncture, modify or disassemble battery.
Do not use the battery that is severely scarred or deformed.
Do not short terminals.
Do not expose to high temperatures.
Do not replace by a battery other than that specified by manufacturer.
Do not charge or discharge otherwise than in the device.
Содержание Volumat Agilia
Страница 1: ...0 9 C E S Volumat Agilia Technical Manual ...
Страница 6: ...66 TM_Volumat_Agilia_Eng_TDM fm ...
Страница 8: ...8 va01_001e_Eng fm Introduction 1 2 Operation diagram ...
Страница 12: ...12 va01_001e_Eng fm Introduction ...
Страница 26: ...26 va02_001e_Eng fm Description and operation ...
Страница 30: ...30 va03 1_001e_Eng fm Description of the menus ...
Страница 62: ...62 va04 4_001e_Eng fm Preventive maintenance ...
Страница 70: ...70 va05 1_001e_Eng fm Troubleshooting ...
Страница 76: ...76 va06 0_001g_Eng fm Intervention procedures ...
Страница 82: ...82 va06 0_0011g_Eng fm Intervention procedures ...
Страница 86: ...86 va06 0_0021g_Eng fm Intervention procedures ...
Страница 94: ...94 va06 0_0040g_Eng fm Fiches d intervention ...
Страница 104: ...104 va06 0_0060g_Eng fm Fiches d intervention ...
Страница 110: ...110 va06 0_0070g_Eng fm Fiches d intervention ...
Страница 115: ...va06 0_0080g_Eng fm 115 Fiches d intervention 12 Remove the left flange marker 12 Pull out the pumping system ...
Страница 134: ...134 va07 1_001g_Eng fm Calibrations ...
Страница 136: ...136 va08 0_001d_Eng fm Spare parts catalogue ...