va07.1_001g_Eng.fm
131
Calibrations
7.1.3 Etal 2, Door
Calibrating the door consists of calibrating the door’s Hall Effect sensor in several phases.
Etal 2
, press "enter".
The main "
Door
" menu displays.
Opened:
open the door.
Press "OK" to validate and move to the next step.
The value is read and recorded.
It should lie between 1400 < X < 1800mV.
Intermediate:
place the calibration tool referenced
Z067059 with the door in the intermediate position.
Press "OK" to validate and move to the next step.
The value is read and recorded.
Closed:
shut the door.
Press "OK" to validate and move to the next step.
The value is read and recorded.
This value should lie between
1500 < X < 2700mV in case the Hall Effect sensor
magnet is mounted in the "south" direction or between 600 < X < 1500 mV in
case the Hall Effect sensor magnet is mounted in the "north" direction.
Press again to confirm and save the calibration values.
If the values are out of range, a message indicating the
non-conformity appears: press "Exit".
Restart the calibration procedure.
Содержание Volumat Agilia
Страница 1: ...0 9 C E S Volumat Agilia Technical Manual ...
Страница 6: ...66 TM_Volumat_Agilia_Eng_TDM fm ...
Страница 8: ...8 va01_001e_Eng fm Introduction 1 2 Operation diagram ...
Страница 12: ...12 va01_001e_Eng fm Introduction ...
Страница 26: ...26 va02_001e_Eng fm Description and operation ...
Страница 30: ...30 va03 1_001e_Eng fm Description of the menus ...
Страница 62: ...62 va04 4_001e_Eng fm Preventive maintenance ...
Страница 70: ...70 va05 1_001e_Eng fm Troubleshooting ...
Страница 76: ...76 va06 0_001g_Eng fm Intervention procedures ...
Страница 82: ...82 va06 0_0011g_Eng fm Intervention procedures ...
Страница 86: ...86 va06 0_0021g_Eng fm Intervention procedures ...
Страница 94: ...94 va06 0_0040g_Eng fm Fiches d intervention ...
Страница 104: ...104 va06 0_0060g_Eng fm Fiches d intervention ...
Страница 110: ...110 va06 0_0070g_Eng fm Fiches d intervention ...
Страница 115: ...va06 0_0080g_Eng fm 115 Fiches d intervention 12 Remove the left flange marker 12 Pull out the pumping system ...
Страница 134: ...134 va07 1_001g_Eng fm Calibrations ...
Страница 136: ...136 va08 0_001d_Eng fm Spare parts catalogue ...