Connect the 50 Ohm load to the open port on the Power splitter and start a calibration sweep by pressing
“Analyze” then “Calibrate”. The calibration sweep takes approximately 1 minute to complete.
Note:
Whenever the Center Frequency, Maximum Length, or Velocity Factor are changed the Cable Fault
instrument will force a recalibration before performing an analysis sweep.
Remove the 50 Ohm load and connect the cable to be tested - choose either Single Sweep or Continuous
in the “Analyze” submenu. Each sweep takes approximately 1 minute to complete and the display will
update with a plot of return loss vs. distance.