212
2.2.8.1
Calibration
The R8100 has a calibration function to assure accurate and consistent performance over time and in a
variety of environmental settings. The options presented by the Calibration soft key are as follows:
Calibrate Now
This initiates a self alignment process for the R8100 to compensate for component aging or adjust
operation to a significant change in external environmental temperature. New calibration values are
generated and stored in memory and used for future operation of the R8100. See Appendix C – R8100
Field Calibration Procedure
Load Factory Defaults
Note: This is an upcoming feature. The calibration values generated during factory alignment prior to
shipment are loaded. These establish a nominal baseline for R8100 operation in the event the user
generated self calibration tables are lost or corrupted.
2.2.8.2
System Settings
The R8100 has hardware configuration settings that are accessed in the Systems Settings submenu. Once
set these parameters aren’t affected by the operational mode or other choices in the R8100 menu system.
Figure 3.2.9.2-1 shows the Systems Settings submenu and the options are as follows:
Содержание R8100 SERIES
Страница 19: ...19 Figure 2 1 2 Left and Right Sides of the R8100...
Страница 164: ...164 Figure 2 2 7 7 5 Power Profile Frame with Markers...
Страница 224: ...224 Figure 3 2 9 6 1 Submenu after pressing Options soft key in the Settings menu...
Страница 243: ...243 PTC Test in Monitor Mode Eye Diagram Display...
Страница 244: ...244 Power Profile Display Linear and Circular Constellation Displays...
Страница 245: ...245 Symbol Distribution Plot...