203
Default: No
RF In/Out (Test Preferences submenu)
Specifies the gain or loss between the RF In/Out port and the unit under test.
In the case of a loss, such as a cable or attenuator, enter a
negative
value. The affected field values will be
increased
to compensate. In Generate mode, for example, if a -6.0 dB value is entered, the Output Level
amplitude is
increased
by 6 dB. In Monitor mode, for example, the input level is
increased
by 6 dB.
In the case of a gain, such as an amplifier, enter a
positive
value. The affected fields will be
decreased
to
compensate. In Generate mode, for example, if a 10.0 dB value is entered, the Output Level amplitude
will
decrease
by 10 dB. In Monitor mode, for example, Wattmeter measurements will be
reduced
by 10
dB.
This value is not used if RF Level Offset is disabled (Off). This field mirrors the RF In/Out field described
in System Settings.
Recommended: Site preference
Default: 0.0
Save Preferences (Test Preferences submenu)
Stores the edited preferences for use by all future activities.
Return (Test Preferences submenu)
Returns to the Main AutoTune menu.
2.2.7.11
AutoScript (BETA) Test Mode
AutoScript provides monitor and control (M&C) script execution from within the R8100. Using this
feature, repetitive measurements normally requiring several keystrokes can be programmed into plain
text scripts and the results written to an exportable log file. Scripts are imported into the R8100 using a
USB drive and may be combined together to form procedures. Procedures can be saved and recalled to
run a particular set of scripts.
This feature is currently in beta testing. Future AutoScript versions may have an altered user interface,
different script format, or other significant modifications as user feedback warrants. Recommended
current use is limited to small-scale R8100 measurement automation only.
Содержание R8100 SERIES
Страница 19: ...19 Figure 2 1 2 Left and Right Sides of the R8100...
Страница 164: ...164 Figure 2 2 7 7 5 Power Profile Frame with Markers...
Страница 224: ...224 Figure 3 2 9 6 1 Submenu after pressing Options soft key in the Settings menu...
Страница 243: ...243 PTC Test in Monitor Mode Eye Diagram Display...
Страница 244: ...244 Power Profile Display Linear and Circular Constellation Displays...
Страница 245: ...245 Symbol Distribution Plot...