Device Functions Proline Promass 83
7 Block OUTPUTS
Function description
OUTPUTS
o
PULSE/FREQUENCY OUTPUT (1...2)
o
CONFIGURATION (FREQUENCY)
ASSIGN FREQUENCY
(4201)
!
Note!
This function is not available unless the FREQUENCY setting was selected in the
OPERATION MODE function (4200).
Use this function to assign a measured variable to the frequency output.
Options:
OFF
MASS FLOW
VOLUME FLOW
CORRECTED VOLUME FLOW
DENSITY
REFERENCE DENSITY
TEMPERATURE
Advanced options with optional software package CONCENTRATION:
TARGET MASS FLOW
% TARGET MASS FLOW
TARGET VOLUME FLOW
% TARGET VOLUME FLOW
CORRECTED TARGET VOLUME FLOW
CARRIER MASS FLOW
% CARRIER MASS FLOW
CARRIER VOLUME FLOW
% CARRIER VOLUME FLOW
CORRECTED CARRIER VOLUME FLOW
% BLACK LIQUOR
°BAUME
°API
°PLATO
°BALLING
°BRIX
OTHERS (_ _ _ _ flexible concentration)
Advanced options with optional software package ADVANCED DIAGNOSIS:
MASS FLOW DEVIATION
DENSITY DEVIATION
REFERENCE DENSITY DEVIATION
TEMPERATURE DEVIATION
TUBE DAMPING DEVIATION
ELECTRODYNAMIC SENSOR DEVIATION
OPERATING FREQUENCY FLUCTUATION DEVIATION
TUBE DAMPING FLUCTUATION DEVIATION
Factory setting:
MASS FLOW
!
Note!
If you select OFF, the only function shown in the CONFIGURATION function group is
this function, in other words ASSIGN FREQUENCY (4201).