Issue/Rev. 0.2 (12/12)
MN0M023 • Page 129
Device Functions Proline Promass 83
9 Block BASIC FUNCTION
9.2.4
Function group ADJUSTMENT
BASIC FUNCTION
G
HART
GAA
PROCESS PARAMETER
GIA
CONFIGURATION
640
EPD PARAMETER
642
REFERENCE PARAMETER 646
ADJUSTMENT
648
Function description
BASIC FUNCTION
o
PROCESS PARAMETER
o
ADJUSTMENT
ZEROPOINT
ADJUSTMENT
(6480)
This function enables a zero point adjustment to be automatically carried out. The new
zero point determined by the measuring system is adopted by the function ZEROPOINT
(see page 135).
Options:
CANCEL
START
Factory setting:
CANCEL
"
Caution!
Before carrying this out, please refer to the Operating Instructions
Proline Promass 83
,
BA 059D/06/en for a detailed description of the procedure for zero point adjustment.
!
Note!
• Programming is locked during zero point adjustment.
The message“ZERO ADJUST RUNNING” appears on the display.
• If the zero point adjustment is not possible (e.g. if v > 0.1 m/s) or has been canceled,
the alarm message “ZERO ADJUST NOT POSSIBLE” appears on the display.
• If the Promass 83 measuring electronics are fitted with a status input, then the zero
point can also be activated by using this input.