Issue/Rev. 0.2 (12/12)
MN0M023 • Page 105
Device Functions Proline Promass 83
7 Block OUTPUTS
7.3.2
Function group OPERATION
OUTPUTS
E
CURRENT OUTPUT 1
EAA
CURRENT OUTPUT 2
EAB
CURRENT OUTPUT 3
EAC
PULSE/FREQ. OUTPUT 1 ECA
PULSE/FREQ. OUTPUT 2 ECB
RELAY OUTPUT 1
EGA
CONFIGURATION
470
OPERATION
474
RELAY OUTPUT 2
EGB
CONFIGURATION
470
OPERATION
474
Function description
OUTPUTS
o
RELAY OUTPUT (1...2)
o
OPERATION
ACTUAL STATUS RELAY
OUTPUT
(4740)
Use this function to check the current status of the relay output.
A jumper on the contact side defines the relay output as a normally open (NO or make) or
normally closed (NC or break) contact (see Operating Instruction
Proline Promass 83
, BA
059D/06/en).
Display:
MAKE CONTACT OPEN
MAKE CONTACT CLOSED
BREAK CONTACT OPEN
BREAK CONTACT CLOSED
SIMULATION SWITCH
POINT
(4741)
Use this function to activate simulation of the relay output.
Options:
OFF
ON
Factory setting:
OFF
!
Note!
• The “SIMULATION RELAY” message indicates that simulation is active.
• The measuring device continues to measure while simulation is in progress, i.e. the
current measured values are output correctly via the other outputs.
• If the “BATCH VALVE 1” option was selected in the ASSIGN RELAY (4700) function,
the functional test takes place by means of the BATCH PROCEDURE (7260) function,
see p. 162.
"
Caution!
The setting is not saved if the power supply fails.