10.4. Command List
The table below describes the available commands via RS485 or LAN/Ethernet interface.
Table 6: Command List
Description
Char Format
(2)
P
(1)
B
(1)
S
(1)
Legal Values
Factory
Default
Burst string format
$
(3)
√
√
(3)
UTSI
Show list of commands
?
√
Ambient correction
A
√
min/max range
Low end of
sensor range
Advanced hold w.
average
AA
√
0.0-300.0s
000.0
Ambient compens.
control
AC
√
0, 1 or 2
0
Top of mA range
AH
√
min/max range
High end of
sensor range
Bottom of mA range
AL
√
min/max range
Low end of
sensor range
Measured attenuation
B
nn - nn
√
√
00 to 99%
n/a
Burst speed
BS
n - nnnnn
√
√
5 - 10000msec
32msec
Advanced hold threshold C
√
min/max range
Low end of
sensor range
Current emissivity
CE
n.nnn
– n.nnn
0.100
– 1.100
1.000
Baud rate (6)
D
nnn - nnnn
√
√
12 = 1200 baud
38400 baud
√
√
24 = 2400 baud
√
√
96 = 9600 baud
√
√
192 = 19200 baud
√
√
384 = 38400 baud
√
√
576 = 57600 baud
√
√
1152 = 115200 baud
Digital filter
DF
n
√
√
0 = OFF, 1 = ON
1
DHCP / BOOTP
DHCP n
√
√
0 = OFF, 1 = DHCP ON
2 = BOOTP ON
0
Sensor gain
DG
n.nnnnnn -
n.nnnnnn
√
0.800000 up to 1.200000
1.000000
Sensor offset
DO
-nnn - +nnn
√
-200 up to +200
0
Emissivity
E
n.nnn
√
√
√
0.100
– 1.100
1.000
Extension board
temperature
EBT
n.n - nnn.n
√
√
0.0
– 999.0 (°C or °F)
Error Codes (9)
EC
nnnnnnnn
√
√
0000
– FFFF (Hex)
Emissivity source
ES
X
√
I or E
I
Valley hold time (4)
F
n.n - nnn.n
√
√
√
0.0
– 300.0 sec (300 s =
)
000.0
Average time (4)
G
n.n - nnn.n
√
√
√
0.0
– 300.0 sec (300 s =
)
000.0
Gateway Address
GW
nnn.nnn.nnn.nnn
√
√
0.0.0.0 - 255.255.255.255 192.168.42.1
Top of mA temperature
range
H
nnnn.n
– nnnn.n √
√
√ min/max range (°C or °F) Upper end of
sensor range
Sensor internal ambient
I
n.n - nnn.n
√
√
0°C/32°F
– 65°C/149°F
Analog input mA
IN
nn.nn
– nn.nn
0-20 or 4-20
Analog input mode
INM
n
0 = 0-20mA, 4 = 4-20mA
4
IP Address
IP
nnn.nnn.nnn.nnn
√
√
0.0.0.1 - 255.255.255.255 192.168.42.132
Switch panel lock
J
X
√
√
L = Locked
U = Unlocked
Unlocked
Relay alarm output
control
K
n
√
√
0 = Permanently Open
2
1 = Permanently Closed
2 = Normally Open
3 = Normally Closed
Содержание endurance series
Страница 4: ......
Страница 8: ...iv...
Страница 10: ...vi...
Страница 24: ...Endurance Series Users Manual 12 3 4 1 1 Measurement Spot Size Diagrams for F0 Narrow Focus Option...
Страница 25: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Technical Data 13...
Страница 26: ...Endurance Series Users Manual 14 3 4 1 2 Measurement Spot Size Diagrams for F1 Close Focus Option...
Страница 27: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Technical Data 15...
Страница 28: ...Endurance Series Users Manual 16 3 4 1 3 Measurement Spot Size Diagrams for F2 Infinity Focus Option...
Страница 69: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Signal Processing 57 defaults are restored...