Endurance® Series
Users Manual
34
6. Device Control
Once you have your sensor(s) positioned and connected properly, the system is ready for
continuous operation. Nonstop operation of the Endurance
®
device is achieved either by back
panel operation or through software control via the RS485, the LAN/Ethernet, PROFINET or
Ethernet/IP communication interface. The Endurance
®
software, a MS-Windows based setup
and configuration program is supplied with your sensor. You can also create custom programs
using the communication protocols listed in section 10, Programming Guide.
6.1. Control Panel
The Endurance
®
fiber optic infrared pyrometers are equipped with a control panel, which is
located inside the sealed electronics housing. It is the user interface for manual operation and
consists of an upper target temperature display, a lower information/configuration display, one
red alarm LED, one green status LED and several setting/controlling buttons, as shown in
Figure 22. The control panel is primarily for setting up the pyrometer prior to nonstop operation.
After the setup and configuration for nonstop operation, the screwable sealed housing cover
has to be closed to protect the user panel for nonstop operation. You are able to configure
sensor settings via the control panel or remotely via a computer or a programmable logic
controller.
Upper Display (green):
7-segment, 4 digits LED type to display the measured
object temperature or error codes.
Lower Display (green/red):
Background illuminated graphics display type. Resolution is
32 * 136 pixels to display 2 text lines of about 16 characters
per line. It is the main screen/menu display, which shows
all information and configuration topics.
LED1 (red):
Indicates the Laser on/off status
LED2 (green):
Sensor alarm status, steady green after warm-up phase
4 Control buttons:
Up, Down, Browse, Enter to step through the menu and for
entering setup values
The sensor has a remote locking feature to protect the unit from accidental interaction over the
control panel. This lockout mode denies access to the submenu functions of the control panel.
Via the RS485, the LAN/Ethernet, the PROFINET IO, the Ethernet/IP communication interface
or a specific key command over the control panel, the Endurance
®
device can be unlocked.
Figure 22: Control Panel
Содержание endurance series
Страница 4: ......
Страница 8: ...iv...
Страница 10: ...vi...
Страница 24: ...Endurance Series Users Manual 12 3 4 1 1 Measurement Spot Size Diagrams for F0 Narrow Focus Option...
Страница 25: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Technical Data 13...
Страница 26: ...Endurance Series Users Manual 14 3 4 1 2 Measurement Spot Size Diagrams for F1 Close Focus Option...
Страница 27: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Technical Data 15...
Страница 28: ...Endurance Series Users Manual 16 3 4 1 3 Measurement Spot Size Diagrams for F2 Infinity Focus Option...
Страница 69: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Signal Processing 57 defaults are restored...