Endurance® Series
Users Manual
8
3.2. Electrical Specifications
Electrical Specifications
Device Model
Parameter
EF1ML, EF1MM, EF1MH,
EF2ML, EF2MH,
EF1RL, EF1RM, EF1RH,
EF2RL, EF2RH
Power Supply
20 to 48 VDC allowed, max. 12W
Power over Ethernet (IEEE 802.3af)
Digital Input (External Trigger / Hold)
Galvanically isolated TTL input signal (digital active low)
To trigger:
-
Average-, Peak-, Valley Hold function
-
Restart of signal postprocessing
-
Switching the optional LASER sighting on/off
Digital Output (Relay / Alarm Output)
Galvanically isolated digital output of a potential-free solid state relay contact.
The max. relay contact load is limited to 48 V, 300 mA. The relay contact
behavior is settable via the user interface or a serial command to operate as
following:
-
NO = Normally Open
-
NC = Normally Close
-
PO = Permanently Open
-
PC = Permanently Close
Analog Input (0/4
– 20mA current loop input)
Galvanically isolated analog input signal, to read in the current (0/4
– 20mA),
given by an external device like PLC, Computer or any kind of process control
device. The internal loop resistor per pyrometer
is about 220Ω.
Via a controlled current from an external device you are able to set/correct:
-
Verification of preset current values, given by external devices
-
Set emissivity (1-Color or 2-Color mode)
-
Set slope (2-Color devices only)
-
Set background temperature for background compensation
Analog Output (current loop, 0/4
– 20mA)
Galvanically isolated analog output signal, to stimulate output currents in the
range (0/4 - 20 mA). The output signal is an active output with 16 Bit
resolution, to drive a maximum resistive load of 500Ω. Just the acquired
object temperature, measured by either 1C or 2C devices, displayed on on
the Endurance® control panel, is converted to a 16 bit current equivalent. The
user has to define the temperature measurement range and must assign the
lower and upper range limits to the current equivalent (0
– 20mA / 4 – 20mA).
RS485 Serial Communication Interface (4-/2-wire full/half duplex transmission)
Galvanically isolated 4- or 2-wire RS485 communication interface. The
ordered 'Electronics Enclosure Configuration' option (M16-connector/Cable
gland) determines the 4-/2-wire capability.
-
Data format: 8 bit, no parity, 1 stop bit
-
Data rate (Bit/s): 1200, 2400, 9600, 19200, 38400, 57600, 115200
-
Default data rate (factory setup): 38400 Bit/s
Digital Network Communication Interface
Galvanically isolated 4-wire network communication interface:
-
Full duplex, 100 Mbit (100Base-TX / IEEE 802.3u)
-
Power over Ethernet (PoE) capability, referred to PoE
standard IEEE 802.3af, mode A, 10/100 Mbit, mixed DC & data
-
TCP/IP, UDP, HTTP, Webserver, MJPEG-Video
-
Additional communication options (protocol stacks) orderable
Содержание endurance series
Страница 4: ......
Страница 8: ...iv...
Страница 10: ...vi...
Страница 24: ...Endurance Series Users Manual 12 3 4 1 1 Measurement Spot Size Diagrams for F0 Narrow Focus Option...
Страница 25: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Technical Data 13...
Страница 26: ...Endurance Series Users Manual 14 3 4 1 2 Measurement Spot Size Diagrams for F1 Close Focus Option...
Страница 27: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Technical Data 15...
Страница 28: ...Endurance Series Users Manual 16 3 4 1 3 Measurement Spot Size Diagrams for F2 Infinity Focus Option...
Страница 69: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Signal Processing 57 defaults are restored...