Innovative High Temperature Fiber Optic Infrared Pyrometers
Device Control
41
Order of subentry appearance:
1. Subentry: MODE
The subentry MODE is just available for 2-Color (ratio) Endurance
®
sensor devices,
where you can force the device to display the measured temperature values in either
one of both modes. With the
▲ ▼ keys, you can toggle between 1–Color and 2–Color.
2. Subentry: TEMP. UNITS
Shows the configured display temperature unit (°C / °F).
With the
▲ ▼ keys, you can toggle between the display temperature units °C or °F.
3. Subentry: RELAY MODE
Shows the configured RELAY MODE of the potential free relay contact.
With the
▲ ▼ keys, you can toggle between the different relay contact behaviors like:
NORMALLY OPEN
PERMANENT CLOSED
PERMANENTLY OPEN
NORMALLY CLOSED
4. Subentry: ATTENUATION RELAY
Shows the configured ATTENUATION RELAY in % of attenuation.
With the
▲ ▼ keys, you can toggle between 0% to 95% of attenuation.
5. Subentry: ATTENUATION FAILSAFE
Shows the configured ATTENUATION FAILSAFE in % of attenuation.
With the
▲ ▼ keys, you can toggle between 0% to 95% of attenuation.
6. Subentry: LASER/LED/CAMERA
Shows the firmware identified pointing device, regarding the ordered sighting option.
If the Endurance
®
firmware cannot identify a pointing device, then NO DEVICE FOUND
will be displayed. If an identified pointing device (LASER, LED, CAMERA) is present,
you can toggle with the
▲ ▼ keys, between ON and OFF to activate or deactivate the
pointing device. After an ON confirmation by ENTER-key, the red pointing device LED
shows the activation status and the high intensity LASER or LED is working.
Do not look direct into the LASER or LED beam, if activated
7. Subentry: FACTORY DEFAULT
Shows, if the Endurance
®
device shall be configured (preset) by factory default values.
With the
▲ ▼ keys, you can toggle between NO and YES.
8. Subentry: KEY -ENTER-
Shows the LOCKED / UNLOCKED status to avoid unintended user control interactions,
if the Endurance
®
device is in permanent network or data transmission process. Via a
serial or network command, the control user interface can be locked or unlocked.
With the
▲ ▼ keys, you can toggle between LOCKED and UNLOCKED, to retrieve
user access by the control interface.
Содержание endurance series
Страница 4: ......
Страница 8: ...iv...
Страница 10: ...vi...
Страница 24: ...Endurance Series Users Manual 12 3 4 1 1 Measurement Spot Size Diagrams for F0 Narrow Focus Option...
Страница 25: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Technical Data 13...
Страница 26: ...Endurance Series Users Manual 14 3 4 1 2 Measurement Spot Size Diagrams for F1 Close Focus Option...
Страница 27: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Technical Data 15...
Страница 28: ...Endurance Series Users Manual 16 3 4 1 3 Measurement Spot Size Diagrams for F2 Infinity Focus Option...
Страница 69: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Signal Processing 57 defaults are restored...