5.2. Mounting the Sensor Head of Laser sighting devices
Sensors equipped with a sighting laser use a branched fiber optic cable. Insert the cable
through the bushing and behind the laser module. Note: one branch is large (sensor) and one
branch is small (sighting laser).
Laser Module
Note keyed connecting sleeve (can be inserted only one way
- slot must face mode switches)
Push large connecting sleeve in until it stops
(approx. 15 mm / 0.6 in)
Tighten screw (finger tighten only)
Tighten only after securing both connecting sleeves
Tighten set screw with 1.3 mm (0.05 in.) hex wrench
Laser aperture: push small connecting sleeve in until it stops
Figure 10: Mounting the Fiber Optic Cable for Laser sighting devices