8. Device Options
Options are items that are factory installed and must be specified at time of order. The
following are available:
Fiber optic cable lengths:
1, 3, 6, 10 m (3, 10, 20, 33 ft), 22 m (72 ft) for selected models
ISO Calibration Certificate, based on NIST/DAkkS certified probes (E-CERT)
High Temperature Fiber Cable (HT), rated to 315°C (600°F)
Laser Sighting (L) for selected models
Communication option (0 = Standard, 1 = PROFINET, 2 = Ethernet/IP)
The High Temperature Fiber Cable excludes PTFE coating and IP65
(NEMA-4) rating!
8.1. Focus (3 focus options available)
In dependence of the scheduled operation and environment, the customer has to select the
right focus distance prior to place the order.
F0 (Narrow Focus) focused at 100mm (4'')
F1 (Close Focus) focused at 300mm (12'')
F2 (Standard Focus) focused
at infinity (∞)
8.2. Laser Sighting
(Sighting Option L)
The laser sighting allows fast and precise aiming at small, rapidly moving targets, or targets
passing at irregular intervals. The laser is specially aligned with the sensor’s lens to provide
accurate, non-parallax pinpointing of targets. The laser comes as a small, bright red spot
indicating the center of the area being measured.
Figure 42: LASER Sighting Indication
To preserve laser longevity, the laser automatically turns off after
approximately 10 minutes of constant use!
Avoid exposure to LASER light! Eye damage can result.
Use extreme caution when operating!
Never look direct into the LASER beam.
Never point directly at another person!
Mirror and dispersion effects can injure Eyes.
Laser dot
Spot marker
Measured spot at the inner of the circle
Содержание endurance series
Страница 4: ......
Страница 8: ...iv...
Страница 10: ...vi...
Страница 24: ...Endurance Series Users Manual 12 3 4 1 1 Measurement Spot Size Diagrams for F0 Narrow Focus Option...
Страница 25: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Technical Data 13...
Страница 26: ...Endurance Series Users Manual 14 3 4 1 2 Measurement Spot Size Diagrams for F1 Close Focus Option...
Страница 27: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Technical Data 15...
Страница 28: ...Endurance Series Users Manual 16 3 4 1 3 Measurement Spot Size Diagrams for F2 Infinity Focus Option...
Страница 69: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Signal Processing 57 defaults are restored...