Innovative High Temperature Fiber Optic Infrared Pyrometers
Installation
33
multidrop manner, whereas all A-signals, as well all B-signals have to be connected to common
lines. The common A-signals have to be routed to the TX+ and the common B-signals to TX-
terminal at the selected USB/RS485 converter.
5.5.6. Addressing the Endurance
®
sensor in a RS485 Multidrop Network
If you are installing two or more sensors in a multi-drop configuration, please be aware of the
following:
Each sensor must have a unique address greater zero (1 - 32).
Each sensor must be set to the same baud rate (default is 38.4 kBaud).
Once all the units are addressed, wire up the units in the 2-wire multidrop manner,
keeping all A & B to be common.
Now you can run the supplied Endurance
®
software, an own written communication
software or an individual terminal program to access the Endurance
®
sensor for issuing
commands and receive the responses.
Содержание endurance series
Страница 4: ......
Страница 8: ...iv...
Страница 10: ...vi...
Страница 24: ...Endurance Series Users Manual 12 3 4 1 1 Measurement Spot Size Diagrams for F0 Narrow Focus Option...
Страница 25: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Technical Data 13...
Страница 26: ...Endurance Series Users Manual 14 3 4 1 2 Measurement Spot Size Diagrams for F1 Close Focus Option...
Страница 27: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Technical Data 15...
Страница 28: ...Endurance Series Users Manual 16 3 4 1 3 Measurement Spot Size Diagrams for F2 Infinity Focus Option...
Страница 69: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Signal Processing 57 defaults are restored...