5.5.1.1. M16 12-Pin DIN Connector Signal Assignment
In case wiring/re-wiring a M16 12-socket DIN connector or a supplied accessory cable
connector, refer to the following illustration and table for the wiring layout.
Figure 14: M16 connector, the corresponding socket and the wire coding table
5.5.1.2. M12 4-Socket LAN/Ethernet Connector
The LAN/Ethernet connector on Endurance
®
-Series side is a M12 4-socket connector type, D-
coded, suited for industrial Ethernet with IP67 protection rate and a screw retention feature. Via
the LAN/Ethernet connector the Endurance
®
-Series device can also be powered as a PD
(Powered Device) by a PSE (Power Sourcing Equipment) in a PoE (Power over Ethernet)
mode. In such operation mode a PoE injector or a PoE switch is needed. Refer to PoE standard
IEEE 802.3af, mode A, 10/100 Mbit mixed DC & data.
Figure 15: M12 Socket (left) and the corresponding cable plug (right)
Pin
Color
Description
A
Black*
A
B
White*
B
C
Grey*
- mA In
D
Purple*
+ mA In
E
White/Drain
Shield
F
Yellow
Trigger
G
Orange
Relay
H
Blue
Relay
J
Green
+ mA Out
K
Brown
– mA Out
L
Black
Power Ground
M
Red
+ 24 VDC
Note: Twisted Pairs*
A/B an A/B and C/D
Signal
Pin RJ45
Pin M12-4
TD+
1
1
TD-
2
3
RD+
3
2
RD-
6
4
Содержание endurance series
Страница 4: ......
Страница 8: ...iv...
Страница 10: ...vi...
Страница 24: ...Endurance Series Users Manual 12 3 4 1 1 Measurement Spot Size Diagrams for F0 Narrow Focus Option...
Страница 25: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Technical Data 13...
Страница 26: ...Endurance Series Users Manual 14 3 4 1 2 Measurement Spot Size Diagrams for F1 Close Focus Option...
Страница 27: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Technical Data 15...
Страница 28: ...Endurance Series Users Manual 16 3 4 1 3 Measurement Spot Size Diagrams for F2 Infinity Focus Option...
Страница 69: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Signal Processing 57 defaults are restored...