Endurance® Series
Users Manual
56
The Reset signal causes a new reading of the current measured temperature and restarts the
selected signal post processing function.
7.12. Factory Defaults
To globally reset t
he unit to its factory default settings, go to the “factory default” menu item
under the configuration screen menu display. The baud rate and communications mode (single
device or multiple devices / multidrop) will not be affected.
Table 4: Factory Defaults
Parameter
E1M, E2M, E3M
E1R, E2R
Mode (1C / 2C)
1C
2C
Temperature Unit (°C / °F)
°C
°C
Slope
n/a
1.000
Emissivity
1.000
1.000
Transmissivity
1.00
1.00
Average
0.0
0.0
Peak Hold
0.0
0.0
Valley Hold
0.0
0.0
SETPOINT in (°C / °F)
0.0
0.0
DEADBAND in (°C / °F)
2
2
RS485 Communication Mode
2-wire , 38.400 Baud *
2-wire, 38.400 Baud *
MULTIDROP ADDRESS
000 (single sensor)
000 (single sensor)
TERMINAL RESISTOR
OFF
OFF
ETHERNET DHCP
OFF
OFF
ETHERNET IP-ADDRESS
192.168.42.132
192.168.42.132
ETHERNET NETMASK
255.255.255.0
255.255.255.0
ETHERNET GATEWAY ADDR.
192.168.42.1
192.168.42.1
ETHERNET PORTNUMBER
6363
6363
WEB SERVER
OFF
OFF
ANALOG OUTPUT MODE
4
– 20mA
4
– 20mA
OUT Lo LIMIT for 4 mA
Low limit sensor temp.
(e.g. 400.0°C)
Low limit sensor temp.
(e.g. 400.0°C)
OUT Hi LIMIT for 20 mA
High limit sensor temp.
(e.g.1800.0°C)
High limit sensor temp.
(e.g.1800.0°C)
ANALOG INPUT MODE
4
– 20mA
4
– 20mA
IN Lo LIMIT for 4 mA
Low limit sensor temp.
(e.g. 400.0°C)
Low limit sensor temp.
(e.g. 400.0°C)
IN Hi LIMIT for 20 mA
High limit sensor temp.
(e.g.1800.0°C)
High limit sensor temp.
(e.g.1800.0°C)
Serial Output Transmission Mode
Burst mode,
Default string = UTSI
Burst mode,
Default string = UTSI
Relay Output Control
Controlled by unit,
NO function,
indicates failsafe alarms
Controlled by unit,
NO function,
indicates failsafe alarms
Set Output Current
Controlled by unit,
4-20 mA
Controlled by unit,
4-20 mA
Lockout Control Panel Access
Unlocked
Unlocked
* RS485 Modes, like Baud Rate or 2-wire half duplex are unchanged, when the factory
Содержание endurance series
Страница 4: ......
Страница 8: ...iv...
Страница 10: ...vi...
Страница 24: ...Endurance Series Users Manual 12 3 4 1 1 Measurement Spot Size Diagrams for F0 Narrow Focus Option...
Страница 25: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Technical Data 13...
Страница 26: ...Endurance Series Users Manual 14 3 4 1 2 Measurement Spot Size Diagrams for F1 Close Focus Option...
Страница 27: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Technical Data 15...
Страница 28: ...Endurance Series Users Manual 16 3 4 1 3 Measurement Spot Size Diagrams for F2 Infinity Focus Option...
Страница 69: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Signal Processing 57 defaults are restored...