
Innovative High Temperature Fiber Optic Infrared Pyrometers
Accessories
65
9.1.1. High Temp. Multi-conductor cable with M16 connector (E-2CCBxx)
Use the High Temp. 12-wire multi conductor cable (E-2CCBxx) for wiring the Endurance
®
sensor with the 24 VDC power supply, all inputs, outputs, and the RS485 interface. It is a
shielded 12-conductor cable, made of 2 twisted pairs plus 8 separate wires, equipped with a
M16 DIN connector on one side and wire sleeves at the counter side. The cable is Teflon coated
and withstands ambient temperatures from -80 to 200°C (-112°F to 392°F). Teflon coated
temperature cables have well to excellent resistance to oxidation, heat, weather, sun, ozone,
flame, water, acid, alkalis, and alcohol, but poor resistance to gasoline, kerosene, and
degreaser solvents.
Purchasable High Temp. 12-wire multi-conductor cables lengths are 4m (13ft.), 8m (26ft.), 15m
(50ft.), 30m (100ft.), 60m (200ft.), see Table 5.
Temperature:
UL-rated at -80 to 200°C (-112°F to 392°F)
Cable material
Teflon
Cable diameter:
7 mm (0.275 in) nominal
Conductors:
Power supply
2 wires (black/red)
Conductor:
0.3 mm² (AWG 22), stranded tinned copper
Isolation:
FEP 0.15 mm wall (0.006 in)
Shield:
none
RS485 interface
2 twisted pairs (black/white and purple/gray)
Conductor:
0,22 mm² (AWG 24), stranded tinned copper
Isolation:
FEP 0.15 mm wall (0.006 in)
Shield:
Aluminized Mylar with drain wire
Outputs and Ground
6 wires (green/brown/blue/orange/yellow/clear)
Conductor:
0,22 mm² (AWG 24), stranded tinned copper
Isolation:
FEP 0.15 mm wall (0.006 in)
Shield:
none
Teflon develops poisonous gasses, when it is exposed to flames!
If you cut the cable to shorten it, notice that both sets of twisted-pair wires
have drain wires inside their insulation. These drain wires (and the white wire
that is not part of the twisted pair) must be connected to the terminal labeled
CLEAR.
If you purchase your own High Temp. Multi-conductor cable, use wire with the
same specifications as herein mentioned. Maximum RS485 cable length is
1.200 m (4000 ft). Power supply (24VDC) feed in distance to the Endurance
®
sensor should not extend the 60m (200 ft.) limit.
An ordered Multi-Conductor Cable does
not
include a terminal block!
Figure 43: High Temp. Multi-Conductor Cable with M16 Connector (E-2CCBxx)
Содержание endurance series
Страница 4: ......
Страница 8: ...iv...
Страница 10: ...vi...
Страница 24: ...Endurance Series Users Manual 12 3 4 1 1 Measurement Spot Size Diagrams for F0 Narrow Focus Option...
Страница 25: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Technical Data 13...
Страница 26: ...Endurance Series Users Manual 14 3 4 1 2 Measurement Spot Size Diagrams for F1 Close Focus Option...
Страница 27: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Technical Data 15...
Страница 28: ...Endurance Series Users Manual 16 3 4 1 3 Measurement Spot Size Diagrams for F2 Infinity Focus Option...
Страница 69: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Signal Processing 57 defaults are restored...