MPS450
Operators Manual
1-16
R-wave Rate
........................................................... 30, 60, 80, 120, 200, and 250 BPM
R-wave Width
........................................................ 20 to 200 ms (10 ms steps)
Additional Steps: 8, 10, and 12 ms
Rate Accuracy
.......................................................
±
1 %
Amplitude Accuracy
.............................................
±
2 %, Lead II (Exception:
±
5 % for R waves
≤
20 ms)
Power-On Default
................................................. 1.0 mV, square wave 2.0 Hz, R-wave rate 60 BPM, R-wave width
10 ms
Respiration
Rate
........................................................................ 0 (OFF), 15, 20, 30, 40, 60, 80, 100, 120 BrPM
Impedance Variations (
Δ
Ω
)
................................. 0.2, 0.5, 1, or 3
Ω
Accuracy Delta
......................................................
±
10 %
Baseline
................................................................. 500, 1000, 1500, 2000
Ω
, Leads I, II, III
Accuracy Baseline
................................................
±
5 %
Respiration Lead
................................................... LA or LL
Apnea Selection
.................................................... 12, 22, or 32 seconds (one-time events), or continuous (Apnea ON =
respiration OFF)
Power-On Default
.................................................. 20 BrPM, delta 1.0
Ω
, 1000-
Ω
baseline, left-arm lead (LA), 12-second
apnea
Blood Pressure
Input/output Impedance
....................................... 300
Ω
±
10 %
Exciter Input Range
.............................................. 2.0 to 16.0 V rms
Exciter-Input Frequency Range
........................... DC to 5000 Hz
Transducer Sensitivity
......................................... 5 or 40
μ
V/V/mmHg
Pressure Accuracy
...............................................
±
(2 % of s 2 mmHg)
Static Levels, p1
.................................................... -10, 0, 80, 160, 240, 320, 400 mmHg
Static Levels, p2
.................................................... -10, 0, 50, 100, 150, 200, 240 mmHg
Static Levels, p3
.................................................... -5, 0, 20, 40, 60, 80, 100 mmHg
Static Levels, p4
.................................................... -5, 0, 20, 40, 60, 80, 100 mmHg
Dynamic Waveforms, p1
...................................... Arterial: 120/80
Radial artery: 120/80
Left ventricle: 120/00
Right ventricle: 25/00
Dynamic Waveforms, p2
...................................... Arterial: 120/80
Radial artery: 120/80
Left ventricle: 120/00
Right atrium (central venous or CVP): 15/10
Right ventricle: 25/00
Pulmonary artery: 25/10
Pulmonary-artery wedge: 10/2
Left atrium: 14/4
Dynamic Waveforms, p3
...................................... Arterial: 120/80
Radial artery: 120/80
Left ventricle: 120/00
Right atrium (central venous or CVP): 15/10
Right ventricle: 25/00
Pulmonary artery: 25/10
Pulmonary-artery wedge: 10/2
Left atrium: 14/4
Dynamic Waveforms, p4
...................................... Swan-Ganz sequence:
Right atrium (CVP)
Right ventricle RV)
Pulmonary artery (PA)
Pulmonary-artery wedge (PAW)
Respiration Artifact
............................................... BP delta changes from 3 to 16 mmHg
BP Output
.............................................................. Mini DIN 7-Pin
Power-On Default
.................................................. 0 mmHg, transducer sensitivity 5 V/V/mmHg
Temperature
Temperature
.......................................................... 0
°
C (42
°
F), 24
°
C (75.2
°
F), 37
°
C (98.6
°
F), and 40
°
C (104
°
F)
Содержание Biomedical MPS450 Operators
Страница 6: ......
Страница 13: ...Contents continued vii List of Figures Figure Title Page 1 1 MPS450 Patient Simulator 1 8...
Страница 14: ...MPS450 Operators Manual viii...
Страница 16: ......
Страница 32: ...MPS450 Operators Manual 1 18...
Страница 74: ...MPS450 Operators Manual 2 42...
Страница 76: ...MPS450 Operators Manual 3 2...
Страница 82: ...MPS450 Operators Manual 3 8...