Cardiac Functions
Blood Pressure Function
2
2-29
gje027.eps
2.
To display the screen for BP channel 3, select
BP3
:
gje028.eps
3.
To reset BP channel 3 to zero, select
ZERO
.
4.
Press the
ESC
key to return to the top menu BLOOD PRESSURE.
5.
For each channel, select its screen from the top menu, and then select
ZERO
, until
the top menu BLOOD PRESSURE displays 0 as the active setting for all four
channels.
Alternatively, in the Numeric-Control Mode, to zero all four BP channels simultaneously,
press the number keys
007
, and select
RUN
.
Setting Static-Pressure Levels
The MPS450 offers seven preconfigured static-pressure-level settings (mmHg) for each
BP channel, with each channel controlled independently:
Set BP channel 1 to -10, 0, 80, 160, 240, 320, or 400.
Set BP channel 2 to -10, 0, 50, 100, 150, 200, or 240.
Set each of BP channels 3 and 4 to -5, 0, 20, 40, 60, 80, or 100.
Action in the Menu-Control Mode
1.
Press the top-menu key labeled
BP
.
2.
Select
BP1, BP2, BP3,
or
BP4
to display the channel’s screen.
3.
Select
STAT
to display the STATIC screen for the selected channel (in this example,
BP channel 1 with an active setting of 80 mmHg):
gje029.eps
4.
On the LCD screen, RUN flashes, because the displayed setting is active.
Содержание Biomedical MPS450 Operators
Страница 6: ......
Страница 13: ...Contents continued vii List of Figures Figure Title Page 1 1 MPS450 Patient Simulator 1 8...
Страница 14: ...MPS450 Operators Manual viii...
Страница 16: ......
Страница 32: ...MPS450 Operators Manual 1 18...
Страница 74: ...MPS450 Operators Manual 2 42...
Страница 76: ...MPS450 Operators Manual 3 2...
Страница 82: ...MPS450 Operators Manual 3 8...