Remote Operations
Codes and Actions
4
4-7
Table 4-6. Numeric Codes for STE Elevation Action (cont.)
Numeric
Control Code
Action
Remote Control
Entry
236
Set ST elevation to -0.7 mV.
STD-0.7
235
Set ST elevation to -0.6 mV.
STD-0.6
234
Set ST elevation to -0.5 mV.
STD-0.5
233
Set ST elevation to -0.4 mV.
STD-0.4
232
Set ST elevation to -0.3 mV.
STD-0.3
231
Set ST elevation to -0.2 mV.
STD-0.2
230
Set ST elevation to -0.1 mV.
STD-0.1
221
Set ST elevation to -0.05 mV.
STD-0.05
220
Set ST elevation to 0.
STD0
219
Set ST elevation to +0.05 mV.
STD+0.05
229
Set ST elevation to +0.1 mV.
STD+0.1
228
Set ST elevation to +0.2 mV.
STD+0.2
227
Set ST elevation to +0.3 mV.
STD+0.3
226
Set ST elevation to +0.4 mV.
STD+0.4
225
Set ST elevation to +0.5 mV.
STD+0.5
224
Set ST elevation to +0.6 mV.
STD+0.6
223
Set ST elevation to +0.7 mV.
STD+0.7
222
Set ST elevation to +0.8 mV.
STD+0.8
ECG Artifact Stimulation
Table 4-7 lists the numeric control codes for the ECG artifact Stimulation functions.
Table 4-7. Numeric Codes for ECG Artifact Stimulation Actions
Numeric
Control Code
Action
Remote Control
Entry
104
Set ECG artifact to off.
EAOFF
105
Set ECG artifact to 50 Hz sine.
EA50
106
Set ECG artifact to 60 Hz sine.
EA60
107
Set ECG artifact to muscle.
EAMSC
108
Set ECG artifact to wandering baseline.
EAWNDR
109
Set ECG artifact to respiration.
EARESP
Содержание Biomedical MPS450 Operators
Страница 6: ......
Страница 13: ...Contents continued vii List of Figures Figure Title Page 1 1 MPS450 Patient Simulator 1 8...
Страница 14: ...MPS450 Operators Manual viii...
Страница 16: ......
Страница 32: ...MPS450 Operators Manual 1 18...
Страница 74: ...MPS450 Operators Manual 2 42...
Страница 76: ...MPS450 Operators Manual 3 2...
Страница 82: ...MPS450 Operators Manual 3 8...