Remote Operations
Codes and Actions
4
4-11
Performance Waves
Table 4-15 lists the numeric control codes for the performance waves functions.
Table 4-15. Numeric Codes for Performance Waves Actions
Numeric
Control Code
Action
Remote Control
Entry
120
Run 2 Hz square performance wave.
SQU2
121
Run .125 Hz square performance wave.
SQU.125
Table 4-15. Numeric Codes for Performance Waves Actions (cont.)
Numeric
Control Code
Action
Remote Control
Entry
122
Run 2 Hz triangle performance wave.
TRI2
123
Run 2.5 Hz triangle performance wave.
TRI2.5
124
Run 30 BPM pulse performance wave.
PUL30
125
Run 60 BPM pulse performance wave.
PUL60
207
Run 0.5 Hz sine performance wave.
SIN0.5
208
Run 5 Hz sine performance wave.
SIN5
209
Run 10 Hz sine performance wave.
SIN10
210
Run 40 Hz sine performance wave.
SIN40
211
Run 50 Hz sine performance wave.
SIN50
212
Run 60 Hz sine performance wave.
SIN60
213
Run 100 Hz sine performance wave.
SIN100
Performance Wave Amplitude
Table 4-16 lists the numeric control codes for the performance wave amplitude functions.
Table 4-16. Numeric Codes for Performance Wave Amplitude Actions
Numeric
Control Code
Action
Remote Control
Entry
272
Set performance amplitude to 0.05 mV.
PFA0.05
273
Set performance amplitude to 0.10 mV.
PFA0.10
274
Set performance amplitude to 0.15 mV.
PFA0.15
275
Set performance amplitude to 0.20 mV.
PFA0.20
276
Set performance amplitude to 0.25 mV.
PFA0.25
277
Set performance amplitude to 0.30 mV.
PFA0.30
278
Set performance amplitude to 0.35 mV.
PFA0.35
279
Set performance amplitude to 0.40 mV.
PFA0.40
280
Set performance amplitude to 0.45 mV.
PFA0.45
Содержание Biomedical MPS450 Operators
Страница 6: ......
Страница 13: ...Contents continued vii List of Figures Figure Title Page 1 1 MPS450 Patient Simulator 1 8...
Страница 14: ...MPS450 Operators Manual viii...
Страница 16: ......
Страница 32: ...MPS450 Operators Manual 1 18...
Страница 74: ...MPS450 Operators Manual 2 42...
Страница 76: ...MPS450 Operators Manual 3 2...
Страница 82: ...MPS450 Operators Manual 3 8...