MPS450
Operators Manual
3-4
Alternatively, in the Numeric-Control Mode, press the number keys for a respiration-lead
setting according to Table 3-1, and select
RUN
:
Table 3-1. Numeric Codes for Respiration-Lead Settings
Respiration-Lead Setting
Numeric Code
RESP LEAD LA
389
RESP LEAD LA STORED
390
RESP LEAD LL
391
RESP LEAD LL STORED
392
For baseline impedance, press the number keys for a respiration-baseline setting
according to Table 3-2, and select
RUN
:
Table 3-2. Numeric Codes for Respiration-Baseline Settings
Respiration-Baseline Setting
Numeric Code
RESP BASE 500 OHM
180
RESP BASE 500 STRD
393
RESP BASE 1000 OHM
181
RESP BASE 1000 STRD
394
RESP BASE 1500 OHM
182
RESP BASE 1500 STRD
395
RESP BASE 2000 OHM
183
RESP BASE 2000 STRD
396
Adjusting the Respiration Rate
Respiration rate is measured by the number of times per minute the lungs inflate
(inspiration) and deflate (expiration), expressed as breaths-per-minute (BrPM). Pediatric
respiration rates are two or three times higher than adult rates. Newborn babies, for
example, breathe from forty to fifty times per minute, while adults typically breathe
approximately fifteen times per minute.
The MPS450 simulates nine preconfigured respiration rates (BrPM): 15, 20, 30, 40, 60,
80, 100, and 120, plus a respiration-off condition.
Action in the Menu-Control Mode
1.
Press the top-menu key labeled
RESP
. If the respiration-rate setting has not been
changed, the screen displays the default setting (20 BRPM):
gje024.eps
Содержание Biomedical MPS450 Operators
Страница 6: ......
Страница 13: ...Contents continued vii List of Figures Figure Title Page 1 1 MPS450 Patient Simulator 1 8...
Страница 14: ...MPS450 Operators Manual viii...
Страница 16: ......
Страница 32: ...MPS450 Operators Manual 1 18...
Страница 74: ...MPS450 Operators Manual 2 42...
Страница 76: ...MPS450 Operators Manual 3 2...
Страница 82: ...MPS450 Operators Manual 3 8...