MPS450
Operators Manual
4-22
Cardiac-Output Wave / Injectate
Table 4-40 lists the numeric control codes for the BP cardiac-output wave / injectate
functions.
Table 4-40. Numeric Codes for Cardiac-Output Wave / Injectate Actions
Numeric
Control Code
Action
Remote Control
Entry
(N/A)
Run CO wave 2.5 l/min.
COW2.5
(N/A)
Run CO wave 5.0 l/min.
COW5.0
(N/A)
Run CO wave 10.0 l/min.
COW10.0
(N/A)
Run CO wave faulty injectate.
COWFLT
(N/A)
Run CO wave left/right shunt.
COWLRS
(N/A)
Run CO wave cal pulse.
COWCAL
(N/A)
Set CO injectate to 0 degrees C.
COI0
(N/A)
Set CO injectate to 24 degrees C.
COI24
090
Turn CO wave off.
CRSET
091
Run CO wave 0 degrees C at 2.5 l/min.
(N/A)
092
Run CO wave 0 degrees C at 5.0 l/min.
(N/A)
093
Run CO wave 0 degrees C at 10.0 l/min.
(N/A)
094
Run CO wave 24 degrees C at 2.5 l/min.
(N/A)
095
Run CO wave 24 degrees C at 5.0 l/min.
(N/A)
096
Run CO wave 24 degrees C at 10.0 l/min.
(N/A)
097
Run CO wave faulty injectate 0 degrees C.
(N/A)
099
Run CO wave faulty injectate 24 degrees C.
(N/A)
098
Run CO wave left/right shunt 0 degrees C.
(N/A)
100
Run CO wave left/right shunt 24 degrees C.
(N/A)
101
Run CO wave cal pulse 0 degrees C.
(N/A)
102
Run CO wave cal pulse 24 degrees C.
(N/A)
FHR Rate (Fixed)
Table 4-41 lists the numeric control codes for the FHR rate (fixed) functions.
Table 4-41. Numeric Codes for FHR Rate (Fixed) Actions
Numeric
Control Code
Action
Remote Control
Entry
409
Run fetal ECG at 60 BPM.
F60
410
Run fetal ECG at 90 BPM.
F90
411
Run fetal ECG at 120 BPM.
F120
412
Run fetal ECG at 140 BPM.
F140
Содержание Biomedical MPS450 Operators
Страница 6: ......
Страница 13: ...Contents continued vii List of Figures Figure Title Page 1 1 MPS450 Patient Simulator 1 8...
Страница 14: ...MPS450 Operators Manual viii...
Страница 16: ......
Страница 32: ...MPS450 Operators Manual 1 18...
Страница 74: ...MPS450 Operators Manual 2 42...
Страница 76: ...MPS450 Operators Manual 3 2...
Страница 82: ...MPS450 Operators Manual 3 8...