Introduction
and
Specifications
Detailed Specifications
1
1-15
Detailed Specifications
Normal-Sinus-Rhythm Waveform
ECG Reference
...................................................... The ECG amplitudes specified are for Lead II (calibration), from the
baseline to the peak of the R wave. All other leads are proportional.
Normal Sinus Rhythm
.......................................... 12-lead configuration with independent outputs referenced to right leg
(RL). Output to 10 Universal ECG Jacks, color-coded to AHA and IEC
Standards.
High-Level Output
................................................. 0.2 V/mV
±
5 % of the ECG amplitude setting available on the BP3
connector.
Amplitude
.............................................................. 0.05 mV to 0.5 mV (0.05 mV steps); 0.5 mV to 5.5 mV (0.5 mV steps)
Amplitude Accuracy
.............................................
±
2 % of setting Lead II
ECG Rate
............................................................... 30, 40, 45, 60, 80, 90, 100, 120, 140, 160, 180, 200, 220, 240, 260,
280, and 300 BPM
Rate Accuracy
.......................................................
±
1 % of setting
ECG Waveform Selection
..................................... Adult (80 ms) or pediatric (40 ms) QRS duration
Superimposed Artifact
......................................... 50 and 60 Hz, muscle, baseline wander, respiration
ST-Segment Elevation
.......................................... Adult mode only. -0.8 mV to +0.8 mV (0.1 mV steps) Additional steps:
+0.05 mV and -0.05 mV
Power-On Default
.................................................. 80 BPM, 1.0 mV, adult QRS, ST-segment elevation of 0 mV, and a
P-R interval of 0.16 seconds
Pacemaker Waveform
Pacer-Pulse Amplitude
......................................... 1, 2, 5, 10 mV
±
10 %
Pacer-Pulse Width
................................................ 0.1, 0.5, 1.0, 1.5, 2.0 ms
±
5 %
Pacing Rate
........................................................... 75 BPM
Waveforms
Atrial
Asynchronous 75 BPM
Demand with frequent sinus beats
Demand with occasional sinus beats
AV sequential
Noncapture (one time)
Nonfunction
Power-On Default
.................................................. Amplitude 5 mV, width 1.0 ms, atrial waveform
Arrhythmia
Baseline NSR
......................................................... 80 BPM
PVC Focus
............................................................. Left focus, standard timing (except where specified)
Supraventricular Arrhythmia
............................... Atrial fibrillation (coarse or fine); atrial flutter; sinus arrhythmia; missed
beat (one time); atrial tachycardia; paroxysmal atrial tachycardia; nodal
rhythm; and supraventricular tachycardia.
Premature Arrhythmia
.......................................... (All one-time events) Premature atrial contraction (PAC); premature
nodal contraction (PNC); PVC1 left ventricular; PVC1 left ventricular,
early; PVC1 left ventricular, R on T; PVC2 right ventricular; PVC2 right
ventricular, early; PVC2 right ventricular, R on T; and multifocal PVCs
Ventricular Arrhythmia
......................................... PVCs 6, 12, or 24 per minute; frequent multifocal PVCs; bigeminy;
trigeminy; multiple PVCs (one-time run of 2, 5, or 11 PVCs); ventricular
tachycardia; ventricular fibrillation (coarse or fine); and asystole
Conduction Defect
................................................ First-, second-, or third-degree heart block; and right- or left-bundle-
branch block
Power-On Default
.................................................. Atrial fibrillation (coarse); PAC; PVCs 6/minute; first-degree heart block
ECG-Performance-Testing
Amplitude
.............................................................. 0.05 to 0.5 mV (0.05 mV steps)
0.5 to 5.5 mV (0.5 mV steps)
Pulse Wave
............................................................ 30, 60 BPM, with 60 ms pulse width
Square Wave
......................................................... 2.0, 0.125 Hz
Triangle Wave
........................................................ 2.0, 2.5 Hz
Sine Wave
.............................................................. 0.5, 5, 10, 40, 50, 60, 100 Hz
R-wave-Detection Waveform
............................... Haver-Triangle
Содержание Biomedical MPS450 Operators
Страница 6: ......
Страница 13: ...Contents continued vii List of Figures Figure Title Page 1 1 MPS450 Patient Simulator 1 8...
Страница 14: ...MPS450 Operators Manual viii...
Страница 16: ......
Страница 32: ...MPS450 Operators Manual 1 18...
Страница 74: ...MPS450 Operators Manual 2 42...
Страница 76: ...MPS450 Operators Manual 3 2...
Страница 82: ...MPS450 Operators Manual 3 8...