Cardiac Functions
ECG Testing
2
2-23
Alternatively, in the Numeric-Control Mode, press the number keys for a performance
wave according to Table 2-17, and select
RUN
:
Table 2-17. Numeric Codes for Wave/Rate Settings
Wave/Rate Setting
Numeric Code
2 HZ SQUARE
120
0.125 HZ SQUARE
121
2.0 HZ TRIANGLE
122
2.5 HZ TRIANGLE
123
30 BPM PULSE
124
60 BPM PULSE
125
0.5 HZ SINE
207
5 HZ SINE
208
10 HZ SINE
209
40 HZ SINE
210
50 HZ SINE
211
60 HZ SINE
212
100 HZ SINE
213
Adjusting Performance-Wave Amplitude
The MPS450 offers a selection of twenty preconfigured settings (mV) for performance-
wave amplitude (on Lead II): 0.05 to 0.50 (in 0.05 steps) and 0.50 to 5.50 (in 0.50 steps).
Action in the Menu-Control Mode
1.
Press the top-menu key labeled
PERF
.
2.
Select
SEL
to toggle to the screen for adjusting performance-wave amplitude
(wave/AMPL), with AMPL in upper-case letters.
3.
Select
DOWN
or
UP
to scroll to the desired performance-wave amplitude.
4.
The amplitude identified on the LCD screen is active when displayed and remains
active until the setting is changed.
Alternatively, in the Numeric-Control Mode, press the number keys for performance-
wave amplitude according to Table 2-18, and select
RUN
:
Table 2-18. Numeric Code for Wave Amplitude Settings
Wave Amplitude Setting
Numeric Code
0.05 mV
272
0.10 mV
273
0.15 mV
274
0.20 mV
275
Содержание Biomedical MPS450 Operators
Страница 6: ......
Страница 13: ...Contents continued vii List of Figures Figure Title Page 1 1 MPS450 Patient Simulator 1 8...
Страница 14: ...MPS450 Operators Manual viii...
Страница 16: ......
Страница 32: ...MPS450 Operators Manual 1 18...
Страница 74: ...MPS450 Operators Manual 2 42...
Страница 76: ...MPS450 Operators Manual 3 2...
Страница 82: ...MPS450 Operators Manual 3 8...