MPS450
Operators Manual
2-32
gje031.eps
Alternatively, in the Numeric-Control Mode, press the number keys for a dynamic-wave
setting according to Table 2-25, and select
RUN
:
Table 2-25. Numeric Codes for Dynamic-Wave Settings
Dynamic-Wave Setting
NumericCode
BP1 Arterial 120/80
060
BP1 Radial Art 120/80
061
BP1 Left Vent 120/00
062
BP1 Right Vent 25/00
063
BP2 Arterial 120/80
070
BP2 Radial Art 120/80
071
BP2 Left Vent 120/00
072
BP2 Right Vent 25/00
073
BP2 Pulmonary Art 25/10
074
BP2 Pulm Art Wedge 10/2
075
BP2 Left Atrium 14/4
076
BP2 Right Atrium CVP 15/10
077
BP3 Arterial 120/80
080
BP3 Radial Art 120/80
081
BP3 Left Vent 120/00
082
BP3 Right Atrium CVP 15/10
083
BP3 Pulmonary Art 25/10
084
BP3 Pulm Art Wedge 10/2
085
BP3 Left Atrium 14/4
086
BP3 Right Vent 25/00
087
Adding Respiration Artifact to the BP Signal
The MPS450 adds respiration artifact to any BP-channel dynamic wave, including the
Swan-Ganz-wave series. The artifact is turned on separately for each of the four BP
channels.
The following peak-to-peak changes (mmHg) are caused by the respiration wave, with
values independent of the programmed respiration amplitude:
for BP channel 1, a delta change from 0 to 16;
Содержание Biomedical MPS450 Operators
Страница 6: ......
Страница 13: ...Contents continued vii List of Figures Figure Title Page 1 1 MPS450 Patient Simulator 1 8...
Страница 14: ...MPS450 Operators Manual viii...
Страница 16: ......
Страница 32: ...MPS450 Operators Manual 1 18...
Страница 74: ...MPS450 Operators Manual 2 42...
Страница 76: ...MPS450 Operators Manual 3 2...
Страница 82: ...MPS450 Operators Manual 3 8...