Remote Operations
Codes and Actions
4
4-17
Table 4-27. Numeric Codes for BP Channel 1: Static-Pressure Levels Actions
Numeric
Control Code
Action
Remote Control
Entry
342
Set BP1 to -10 mmHg static.
P1S-10
343
Set BP1 to zero.
P1S0
344
Set BP1 to 80 mmHg static.
P1S80
345
Set BP1 to 160 mmHg static.
P1S160
346
Set BP1 to 240 mmHg static.
P1S240
347
Set BP1 to 320 mmHg static.
P1S320
348
Set BP1 to 400 mmHg static.
P1S400
BP Channel 2: Static-Pressure Levels
Table 4-28 lists the numeric control codes for the BP channel 2: static-pressure levels
functions.
Table 4-28. Numeric Codes for BP Channel 2: Static-Pressure Levels Actions
Numeric
Control Code
Action
Remote Control
Entry
351
Set BP2 to -10 mmHg static.
P2S-10
352
Set BP2 to zero.
P2S0
353
Set BP2 to 50 mmHg static.
P2S50
354
Set BP2 to 100 mmHg static.
P2S100
355
Set BP2 to 150 mmHg static.
P2S150
356
Set BP2 to 200 mmHg static.
P2S200
357
Set BP2 to 240 mmHg static.
P2S240
BP Channel 3: Static-Pressure Levels
Table 4-29 lists the numeric control codes for the BP channel 3: static-pressure levels
functions.
Table 4-29. Numeric Codes for BP Channel 3: Static-Pressure Levels Actions
Numeric
Control Code
Action
Remote Control
Entry
360
Set BP3 to -5 mmHg static.
P3S-5
361
Set BP3 to zero.
P3S0
362
Set BP3 to 20 mmHg static.
P3S20
363
Set BP3 to 40 mmHg static.
P3S40
Содержание Biomedical MPS450 Operators
Страница 6: ......
Страница 13: ...Contents continued vii List of Figures Figure Title Page 1 1 MPS450 Patient Simulator 1 8...
Страница 14: ...MPS450 Operators Manual viii...
Страница 16: ......
Страница 32: ...MPS450 Operators Manual 1 18...
Страница 74: ...MPS450 Operators Manual 2 42...
Страница 76: ...MPS450 Operators Manual 3 2...
Страница 82: ...MPS450 Operators Manual 3 8...